T
+44 (0)1709 377172
F
+44 (0)1709 377173
E
www.harvesthealthcare.co.uk
28
29
TROUBLESHOOTING
9
CARE, CLEANING & DISINFECTION
Clean and disinfect the bed before placing into service and before each re-use. To clean the
care bed, wipe the bed by hand with a damp cloth. Use suitable cleaning and conditioning
agents for wooden and synthetic furniture.
Household cleaners without ammonium or scouring agents are allowed, but should be
dermatologically tested.
Solvents and scouring agents are not permitted as they damage the various surfaces of the
care bed.
To disinfect: In the homepage of the Robert Koch Institute < http://www.rki.de > you will
find a list dated of 31.05.2007 of approved and generally accepted disinfection agents and
treatments and their use.
Before cleaning and disinfection the mains plug must be unplugged
and hung up safely. Plugs for the handset and the motors which are
inserted in the control unit MUST remain plugged in. This is necessary
to ensure water does not enter the control system.
The brakes must only be released to move the bed. Refer to the Safety
Information
10 TROUBLESHOOTING
Mains plug not plugged in
Insert mains plug into mains
socket
Locking function on handset
activated
Unlock handset
Handset not plugged in
Insert handset into mattress base
motor
Motor unit not plugged in
Plug motor unit into mattress base
motor
No Response
Connecting cables on the
connectors transposed
Check plugs and connectors and
change over plugging in locations
Adjustment functions transposed
9V block battery is discharged
Replace 9V block battery
No function after power failure
Bed only adjusted via the battery.
Mains plug not plugged in
Plug in mains plug and replace the
9V block battery as a precaution
Bed only moves very slowly
FAULT
POSSIBLE CAUSE
SOLUTION
SERVICING
11 SERVICING
111.1 PRINCIPLES
Operators of care beds are obliged according to MPBetreibV (Operators of Medical Products
Ordinance) §4 to guarantee the safe condition of the medical product over their entire service
life.
The test according to the regulation EN 62353 contains the following minimum requirements:
• Visual check
• Measurement of leakage resistance
• Measurement of leakage current
• Functional test
• Overall evaluation
The service life of the care bed depends essentially upon the handling and servicing.
To guarantee safe operation, a visual and functional test including an electrical test must be
carried out at least once a year. For this purpose, proceed according to the technical safety
checklist as per regulation EN 62353 in section 10.2
If you have any doubts about the safety or functionality of the bed or
even a part of the bed as a result of the work performed below, the
bed should under no circumstances be put into service again.
Contact the supplier or manufacturer in this case.