Harvest Healthcare HARVEST 3 PUMP General User/ Safety Manual Download Page 1

E

 [email protected]      www.harvesthealthcare.co.uk

1

PUMP

www.harvesthealthcare.co.uk

General User/Safety Guide

HARVEST 3 PUMP

Summary of Contents for HARVEST 3 PUMP

Page 1: ...E sales harvesthealthcare co uk www harvesthealthcare co uk 1 PUMP www harvesthealthcare co uk General User Safety Guide HARVEST 3 PUMP ...

Page 2: ...T 44 0 1709 377172 F 44 0 1709 377173 2 ...

Page 3: ...2 SAFETY INFORMATION FOR THE OPERATOR 3 3 SAFETY INFORMATION FOR THE USER 3 3 SAFETY INFORMATION FOR THE USER 3 4 SYMBOLS USED 3 4 SYMBOLS USED 3 5 CLEANING DISINFECTION 3 6 SERVICING MAINTENANCE 3 7 SERVICE LIFE DISPOSAL TECHNICAL SPECIFICATION TECHNICAL SPECIFICATION OVERVIEW OVERVIEW OPERATION OPERATION CLEANING CARE CLEANING CARE ROUTINE MAINTENANCE SERVICING ROUTINE MAINTENANCE SERVICING TROU...

Page 4: ...as the effectiveness of the therapy may be reduced All hoses must be free of kinks twists and must be properly connected and positioned so as not to cause any obstruction Do not position the pump in a way that prevents access to the disconnection device mains power plug Ensure the mains lead or pump cannot become trapped or crushed e g by raising or lowering of bed or bed rails or any other moving...

Page 5: ...instructions can be found on page 15 Suitable for continuous use Do not modify the mattress cushion or pump unit in any way Do not connect to any other medical device or equipment Not for use in an oxygen enriched environment Not for use in an outdoor environment Store the system in a clean and dry environment out of direct sunlight Electrical equipment can be hazardous Only authorised technical p...

Page 6: ... and four choices of cycle time 10 15 20 and 25 minutes Harvest Healthcare supply a pump and mattress but it is the user s responsibility to ensure each system meets the guidelines set out in the pump and mattress guide shown below Failure to do this could put the service user at risk This could prolong the recovery process or increase the risk of further tissue damage For information on our mattr...

Page 7: ... This includes for example Incorrect installation Operation by persons who have not been instructed in its use Using the system with non approved parts accessories Using the system if any of the components are damaged or faulty GENERAL INFORMATION 1 DEFINITION OF THE GROUPS MENTIONED OPERATOR An operator is any natural or legal person who uses the equipment or on whose instruction it is used e g n...

Page 8: ...his instruction manual instruct each user in the safe operation of this system before it is put into service for the first time Advise the user of any hazards which may occur if not handled correctly Only persons who have been properly instructed may operate this system This also applies for persons who only operate the system on a temporary basis 3 3 SAFETY INFORMATION FOR THE USER Ensure that th...

Page 9: ... according to EN 60601 1 Handle with care This way up Keep dry Recycling symbol Refers to packaging that can be recycled cardboard Fragile handle with care This product must be disposed of in a designated refuse bin for waste electronic devices WEEE in the European Union Do not dispose of as normal domestic waste No smoking No naked flames Read instructions consult manufacturers guide This symbol ...

Page 10: ...ak in use Any defects damage or signs of wear must be rectified without delay Only original spare parts from Harvest Healthcare may be used otherwise all guarantees or warranties will be invalidated See pages 16 17 for Routine Maintenance and Routine Servicing 3 7 SERVICE LIFE DISPOSAL Our full terms conditions including product warranties are available by request or can be found on our website ww...

Page 11: ... Flow Output Pressure Setting HPU3 10 15 20 25 minutes 340 x 125 x 244mm 3 5 kg 12 lpm Manual PUMP POWER REQUIREMENTS Power Rating Voltage Fuse Medical Classification Safety Standards 13W AC 230V 50Hz T 1AH 250V Type B Applied Part EN 60601 1 EN 60601 1 2 For any further assistance please contact a member of the Harvest Servicing team at servicing harvesthealthcare co uk ...

Page 12: ...after use J H Feed Tube Connector Pipes from the mattress connect here The pump must be able to be disconnected in an emergency for CPR to be administered A Power Switch The power switch is on the right hand side of the pump When pressed to the ON position the switch will illuminate to indicate that there is a power supply to the pump With the switch in the ON position press the power button on th...

Page 13: ...ove pressure relief for the service user F Static Mode The Static Mode can be operated by the user to inflate all the cells for patient care and transfers Press the Static Mode button and the LED will illuminate orange The Static Mode will stop working after 20 mins and the pump reverts back to normal alternating mode G OVERVIEW Bed Hooks Back of Pump Attach the pump to the foot end of the bed Ada...

Page 14: ...rnating action stops in this mode TRANSPORT MODE To turn off the system press the power button to OFF and unplug from the mains supply When the mattress is fully inflated set the pump to the approximate weight of the service user Position the service user in the lying position If required adjust the weight setting to meet the patient s comfort needs Ensure that the comfort setting is altered corre...

Page 15: ...ve blood or other body fluids mattresses should be thoroughly rinsed with clean water to remove any residues This will help prevent any long term compatibility issues associated with disinfectant residues Alternatively disinfection may be achieved by laundering at temperatures not exceeding 80 C for 10 minutes which may include a chlorine rinse TRANSPORT STORAGE Storage conditions as follows 15 C ...

Page 16: ...ividual cells can be wiped clean with a mild antiseptic solution All cells are replaceable and can be sourced from Harvest Healthcare POWER UNIT Disconnect the power unit from the electricity supply before carrying out maintenance repairs or cleaning Check all electrical connections and power lead for signs of wear and damage The power unit can be wiped down with a detergent or disinfectant soluti...

Page 17: ...he Harvest 3 pump should be serviced every 12 months by Harvest Healthcare approved personnel using genuine Harvest Healthcare spare parts SERVICING YOUR SYSTEM Failure to follow the service schedule may invalidate future warranty claims Guarantees Warranties can be found on page 23 ...

Page 18: ...mp is connected to is working correctly 1 2 3 4 Connect the pump to the nearest working mains outlet Replace the plug fuse with the correct 5A fuse Try a different device in the mains outlet 1 2 3 IF PROBLEM PERSISTS Contact Harvest Healthcare technical support Before calling Please ensure you have the serial number and model of equipment Please record details of the results of the recommended tes...

Page 19: ...tes pages are provided at the back of this user manual PLEASE NOTE Inflation can take up to 30 minutes The hoses are routed correctly not kinked and connected to the pump correctly The CPR valve is not trapped and is in the closed position 1 2 There are no leaks in the mattress The tubes in the mattress are not disconnected or kinked 1 2 Disconnect and then re connect the hoses to the outlet on th...

Page 20: ...ppear to be alternating Check that the static mode is not turned on Carefully mark one of the inflated cells with a pen Ensure that there are no kinks in the pipework down the side of the mattress Check the cycle time on the control panel Recommended setting 10 minutes unless a health professional has recommended otherwise 1 2 3 4 Monitor the cell for 7 minutes to see if it deflates Straighten out...

Page 21: ...ss is uncomfortable Check the comfort setting on the pump Set the pump to the correct setting using the guide on the front of the pump case Contact Harvest Healthcare technical support Before calling Please ensure you have the serial number and model of equipment Please record details of the results of the recommended tests Notes pages are provided at the back of this user manual E sales harvesthe...

Page 22: ...709 377173 22 PARTS LIST APPLIED PARTS REPLACEMENT PARTS PU301 PU302 PU303 PU304 PCB Board Front Panel Sticker Case Bed Hooks HPU3 Harvest 3 Pump only PU300 Connector UN012 Filter PU101 PU102 Compressor Bellows PU305 Timer Motor ...

Page 23: ...de user damage That Harvest Healthcare s cleaning disinfecting guidelines have been followed Harvest Healthcare s maintenance guidelines have been followed please refer to the product manual That ALL maintenance has been carried out by a suitably qualified and competent person That all parts used are OEM Original Equipment Manufacturer parts and were supplied by Harvest Healthcare either directly ...

Page 24: ...omplies with all applicable requirements in Appendix I of the EU directive 93 42E EC The following compliance evaluation process was applied Annex VII EU legislation for this type of product covers all safety health and environmental requirements Modifying this product without consultation with Harvest Healthcare Ltd will invalidate this declaration of conformity This declaration of conformity is ...

Page 25: ...E sales harvesthealthcare co uk www harvesthealthcare co uk 25 NOTES ...

Page 26: ...T 44 0 1709 377172 F 44 0 1709 377173 26 NOTES ...

Page 27: ...E sales harvesthealthcare co uk www harvesthealthcare co uk 27 NOTES ...

Page 28: ...care co uk Copyright Harvest Healthcare Harvest Healthcare Limited Company No 07210261 Our Full Terms Conditions are available by request or can be found on our website Harvest Healthcare reserves the right to alter or amend this document without notice Serial No DOCUMENT REFERENCE HPU3 Harvest 3 Pump March 2020 The pump is supplied by Harvest Healthcare Ltd and complies to the standard EN 60601 1...