T
+44 (0)1709 377172
F
+44 (0)1709 377173
E
www.harvesthealthcare.co.uk
18
19
OPERATION
SWITCHING OFF THE SYSTEM
A transport cap is provided to hold the air in the
mattress if the pump is disconnected temporarily
for any reason.
To prevent deflation please ensure that the
transport cap (attached to the air tubes) is fitted
to the 3 pipe connector on the mattress (see
picture)
The alternating action stops in this mode.
TRANSPORT MODE
WEIGHT KG
25 - 75 kg
76 - 100 kg
100 - 254 kg
WEIGHT LB
55 - 167 lb
166 - 220 lb
220 - 560 lb
WEIGHT STONES LB
3.9 - 11 stone 11 lb
12 - 15 stone 10 lb
15 stone 10 lb - 40 stone
LOW
MEDIUM
HIGH
COMFORT CONTROL GUIDE
Turn off and unplug from the mains supply (the battery backup will operate).
Press the function button and hold continuously until the unit has turned off.
(approx. 30 seconds).
1
2
OPERATION
INFLATING THE MATTRESS
Switch on: The pump will automatically begin to inflate and start a 30 minute mattress
inflation mode until the correct pressure is reached.
During this stage
select a weight setting
to suit the service user - Low, Medium or
High. The weight setting is selected by pressing the
Function
button.
1
2
Set the pump to run at the correct pressure to suit the weight of the
service user. Refer to the Comfort Control Guide (see next page) or on
the front of the pump.
DEFLATING THE MATTRESS / CPR VALVE
To deflate the mattress, simply open
the CPR valve and disconnect the
pump from the mattress using the
quick release connector.
IN THE EVENT OF A POWER FAILURE
Should the power supply to the pump be cut, the system will automatically switch to
battery back-up.
If the power is restored, the pump will continue on the selected setting.
The battery will opperate for approx 6-8 hours, afterwhich the system will deactivate
and will therefore need to be connected again to a live mains power supply.
1
2
3
If rapid deflation is required, simply twist the CPR to the
OPEN
position.