4.9 Dual and Quadruple Collect Assembly
Fig. 4.9.1
Setting up
Holder Quadruple
Cannulae
Fig. 4.9.2
Setting up
Quadruple Cannulae for
Tubing
Fig. 4.9.3
Setting up Quadruple
Cannulae for Septa
4.9.1
Place the Holder for Dual or Quadruple Cannulae (1) on the Supporting
Arm so that the Multi Holders for cannula (2) fit into the left slot (3) and the two
holes (3). Secure the holder with the two accompanying screws (4). See
Figure 4.9.1
4.9.2
Collecting in vials with tubing:
Insert the Cannulae for Tubing in the Multi Holders for Cannula (2), adjust the
upper ends of the metal cannulae to the same level as the top of the Plastic
Locker Knobs (5) when they are mounted in the Multi Holders for Cannula.
Secure the cannulae with the screws in the Multi Holders for Cannula (6). See
Figure 4.9.2
4.9.3
Collecting in vials with septa:
Place waste vials in the cassette and move them to the collection position.
Insert the Cannulae for Septa (7) in the Multi Holders for Cannula (2). Adjust
the lower ends of the cannulae into the bottom of the vials (8). Secure the
cannulae with the screws in the Multi Holders for Cannula (6). See Figure
4.9.3
4.9.4
To remove the Holder for Dual or Quadruple Cannulae, untight the two
screws, turn the Multi Holders for cannula (2) so the opening of the locking
washers (9), seen under the supporting arm, are facing you. Lift up the Holder
for Dual or Quadruple Cannulae from the Supporting Arm. See Figure 4.9.1
WARNING: Risk for cannulae damage.
When changing from quadruple mode to single or dual mode, always
change from the
Holder for Quadruple Cannulae
to the proper one
needed for the application.
Do not excessive tight
the screw (6) to not damage the cannula
lumen.
A too loose screw (6) can’t hold the cannula in position.