Publication 5419-013 REV1.0
44
POLL
Sets or displays the polling mode state.
When polling mode is off, prompts are displayed when an event
happens, such as a target being reached.
When polling mode is on, prompts are not displayed when an
event happens, and an XON character is output when the pump is
ready for another command.
When polling mode is remote, the following occurs:
• Prompts are not displayed
• Echo is forced off and the echo command is illegal
Command format:
poll[<sp>[on|off|remote]]
Default state = off
Query response:
<lf>[<addr>]Off
<lf>[<addr>]<prompt> or:
<lf>[<addr>]On
<lf>[<addr>]<prompt><xon> or:
[<addr>] Remote
Examples:
poll
poll<sp>off
RSAVE
Turns On or Off saving rate changes to memory. Without the
On/Off parameter, the current setting is displayed. With the
parameter, continually writing rate changes to memory is turned
on or off accordingly.
This command should be used when sending commands quickly
to avoid damage to the pump memory.
Command format:
rsave[<sp>{On|Off}]
Query response:
<lf>[<addr>]{On|Off}
<lf>[<addr>]<prompt>
Pump Chain Commands
TIME
Sets or displays the date and time.
Command format:
time[<sp>{mm/dd/yy|dd/mm/yy|yy/mm/dd}{<sp>}
{hh:mm:ss}]
Query response:
<lf>[<addr>]##/##/## ## ##:##:## #M
<lf>[<addr>]<prompt>
The date order format is dependent on the settings made when
setting the date from the touch screen. The time format is 24
hours.
Example 1 (display current date and time):
time
Example 2 (set date and time):
time<sp>01/02/16<sp>13:14:15
Note:
If date format has been set to MM/DD/YY, this command sets
the time and date to 13:14:15 (01:14:15 PM) on January 2,
2016.
If date format has been set to DD/MM/YY, this command sets
the time and date to 13:14:15 (01:14:15 PM) on February 1,
2016.
If date format has been set to YY/MM/DD, this command sets
the time and date to 13:14:15 (01:14:15 PM) on February 16,
2001.