Harvard Apparatus Pump 11 Elite OEM Module Series 

Publication 5420-007 Rev B

www.harvardapparatus.com

phone

 

800.272.2775

email

 

[email protected]

Pump 11 Elite OEM 

Module Series

USER’S MANUAL

Summary of Contents for Pico Plus Elite OEM

Page 1: ... Apparatus Pump 11 Elite OEM Module Series Publication 5420 007 Rev B www h ar var d ap p a ra tu s c o m phone 800 272 2775 email bioscience harvardapparatus com Pump 11 Elite OEM Module Series USER S MANUAL ...

Page 2: ... Pump Control 11 Setting up HyperTerminal 11 Pump Chain Commands 12 Using the Pump 11 Elite Series Pump Chain Commands 12 Error Messages 13 Command Errors 13 Argument Errors 13 System Commands 13 Run Commands 15 Rate Commands 15 Volume Commands 16 Time Commands 17 Digital I O Commands 18 Internal Commands 18 Serial Command Notes 18 Pump 11 Elite OEM Module USER S MANUAL phone 508 893 8999 fax 508 ...

Page 3: ...erface 23 RS 485 25 Legacy RJ 11 Option 26 Appendix D Maintenance 27 Maintenance 27 Upgrading Pump 11 Elite Series Software 28 Troubleshooting 29 Appendix E Ordering Information 30 SUBJECT PAGE NO Pump 11 Elite OEM Module USER S MANUAL phone 508 893 8999 fax 508 429 5732 e mail bioscience harvardapparatus com web www harvardapparatus com Pump 11 Elite OEM Module Series User s Manual Publication 54...

Page 4: ...minals of the product ensure that the product is properly grounded MAKE PROPER CONNECTIONS Make sure all connections are made properly and securely Any signal wire connections to the unit must be no longer than 3 meters OBSERVE ALL TERMINAL RATINGS Review the operating manual to learn the ratings on all connections AVOID EXPOSED CIRCUITRY Do not touch any electronic circuitry inside of the product...

Page 5: ...er Hill Road Holliston Massachusetts 01746 using out toll free number 1 800 272 2775 or outside the US call 1 508 893 8999 Email address is bioscience harvardapparatus com Goods will not be accepted for return unless an RMA returned materials authorization number has been issued by our customer service department The customer is responsible for shipping charges for non warranty repairs Please allo...

Page 6: ...e Selection Drive Motor 0 9 Stepper Motor 0 9 Stepper Motor Motor Drive Control Microprocessor with 1 16 microstepping Microprocessor with 1 16 microstepping Number of Microsteps per one rev of Lead Screw 15 360 20 480 Step Resolution 0 069 µm µstep 0 031 µm µstep Step Rate Minimum 27 5 sec µstep 27 5 sec µstep Maximum 26 µsec µstep 26 µsec µstep Pusher Travel Rate Minimum 0 15 µm min 0 068 µm min...

Page 7: ...Operation Continuous Continuous Classification Class I Class I Pollution Degree 1 1 Installation Category II II Supplier Name Harvard Apparatus Harvard Apparatus Supplier Address 84 October Hill Rd Holliston MA 01746 84 October Hill Rd Holliston MA 01746 Supplier Phone Number 508 893 8999 508 893 8999 Regulatory Certifications CE ETL UL CSA WEEE EU RoHS CB Scheme CE ETL UL CSA WEEE EU RoHS CB Sche...

Page 8: ...lus Elite A wide dynamic flow range from 1 26 pl min to 88 40 ml min 0 54 pl min to 11 70 ml min for the Pico Plus Elite can be programmed into the pump Flow rates are selectable with user selected engineering units from ml µl nl pl and hours minutes and seconds Up to 16 kg 35 lbs of linear force is available This force is user adjustable to ensure the right force is applied for the various syring...

Page 9: ...S ON A COMPUTER DAMAGE MAY OCCUR TO THE PUMP OR COMPUTER I O Connector Foot Switch Main Switch DC Power Input Power Source When powering a pump in a chain through an RS 485 cable toggle switch to RS 485 Only one pump can be powered from master through RS 485 RS 485 Connection Pump to Pump USB Serial Input Type B phone 508 893 8999 fax 508 429 5732 e mail bioscience harvardapparatus com web www har...

Page 10: ...ase button 1 and move the pusher block 2 to the approximate length of the syringe 2 Unscrew the bracket clamping knobs 7 8 to loosen the retaining brackets 4 5 on the pusher block 2 and the Syringe holder block 3 3 Unscrew the syringe barrel clamp by unscrewing the syringe barrel clamp knob 6 4 Place the syringe on the syringe holder block 3 Ensure the barrel flange is with in the syringe holder s...

Page 11: ...own list Click OK If the Virtual Comm Port is not known use Device Manager to find it Instructions on setting up the Virtual CommPort Driver are supplied in Appendix E 5 Verify the Settings are as shown below 6 Choose ASCII Setup and select Echo typed characters locally then click OK to complete the setup Pump 11 Elite OEM Module USER S MANUAL phone 508 893 8999 fax 508 429 5732 e mail bioscience ...

Page 12: ... the pump address is not zero prompt Prompt see below The following prompts are returned after a command is executed The pump is idle The pump is infusing The pump is withdrawing The pump stalled T The target was reached If the pump is in poll REMOTE mode there will be no prompts or carraige returns If the pump is in poll ON mode a XON character is added after the prompt If the pump address is non...

Page 13: ...ets or displays the pump address Valid range is 0 to 99 Command format address 0 99 Query response lf Pump address is cr lf prompt Notes Pumps with an address of 0 are masters and pumps with an address between 1 and 99 are slaves BAUD Sets or displays the baud rate Valid baud rates are 9600 19200 38400 57600 115200 128000 230400 256000 460800 and 921600 Note that some computers may not be able to ...

Page 14: ...draw qs iw cr lf prompt or lf Quick Start Withdraw Infuse qs wi cr lf prompt or lf method name cr lf prompt METRICS Displays the pump metrics The operation of this command is covered in the 11 Elite Configuration document Command format metrics POLL Sets or displays the polling mode state When polling mode is off prompts are displayed when an event happens such as a target being reached When polli...

Page 15: ...lf volume unit cr this line is repeated for each syringe size lf prompt VER Displays the short version string Command format ver Query response lf 11 Elite cr lf prompt VERSION Displays the full version string Command format version Query response lf Firmware v cr lf Pump address cr lf Serial number C cr lf Device ID cr lf prompt Run commands IRUN Runs the pump in the infuse direction Quick Start ...

Page 16: ...esponse lf Ramp not set up cr lf prompt or lf xl xxx to xl xxx in seconds cr lf prompt IRATE Sets or displays the infusion rate Quick Start mode only The rate argument may be replaced by max or min to set the maximum or minimum rate respectively lim may be used to display the range limits Command format irate max min lim rate rate units Query response lf xl xxx cr lf prompt or lf xl xxx to xl xxx ...

Page 17: ... prompt SVOLUME Sets or displays syringe volume Quick Start mode only Command format svolume Query response lf ul cr lf prompt or lf ml cr lf prompt TVOLUME Sets or displays the target volume Quick Start mode only Command format tvolume target volume volume units Query response lf Target volume not set cr lf prompt or lf xl cr lf prompt WVOLUME Displays the withdrawn volume Quick Start mode only C...

Page 18: ...and terminated by a carriage return linefeed pair The first integer is the current rate in femtoliters per second The second integer is the infuse time in milliseconds The time has a granularity of 1 millisecond The third integer is the infused volume in femtoliters All three values are for the current direction The flag field consists of six flags Flag one is the motor direction and will be i or ...

Page 19: ... 14 427 20 19 05 30 21 59 50 26 594 60 26 594 Terumo Size Diameter 1 ml vc 6 50 mm 1 ml tb 4 70 mm 2 5 9 5 13 10 15 8 20 20 15 30 23 2 60 29 2 SGE Scientific Glass Engineering Size Diameter 5 µl 0 343 mm 10 0 485 25 0 728 50 1 03 100 1 457 250 2 303 500 3 257 1 ml 4 606 mm 2 5 7 284 5 10 301 10 14 567 25 23 50 27 5 Hamilton Glass All Types Size Diameter 0 5 µl 0 103 mm 1 0 146 2 0 206 5 0 343 10 0...

Page 20: ... 14 160 pl min 14 710 µl min 10 µl 0 485 mm 28 260 pl min 29 400 µl min 25 µl 0 729 mm 63 960 pl min 66 430 µl min 50 µl 1 030 mm 127 700 pl min 132 600 µl min 100 µl 1 457 mm 255 500 pl min 265 400 µl min 250 µl 2 304 mm 638 900 pl min 663 500 µl min 500 µl 3 256 mm 1 276 nl min 1 325 ml min 1000 µl 4 608 mm 2 556 nl min 2 654 ml min 1 ml 4 699 mm 2 658 nl min 2 760 ml min 3 ml 8 585 mm 8 871 nl ...

Page 21: ...596 5 nl min 1 µl 0 1457 mm 1 14 pl min 1 194 µl min 2 µl 0 206 mm 2 28 pl min 2 386 µl min 5 µl 0 343 mm 6 36 pl min 6 615 µl min 10 µl 0 485 mm 12 72 pl min 13 230 µl min 25 µl 0 729 mm 28 74 pl min 29 880 µl min 50 µl 1 030 mm 57 42 pl min 59 650 µl min 100 µl 1 457 mm 114 9 pl min 119 4 µl min 250 µl 2 304 mm 287 4 pl min 298 5 µl min 500 µl 3 256 mm 574 0 pl min 596 1 µl min 1000 µl 4 608 mm ...

Page 22: ...Default Low Control thru Method or serial comm 9 13 Signal Return Ground 14 Run Indicator Voltage 5V through a 470 ohm resistor connect to LED Anode 15 Run Indicator Output TTL Logic Output active low low run connect to LED Cathode Electrical Specifications Inputs VIH 2V IIH 20µA VIL 0 4V IIL 0 5mA Outputs VOH 3 8V IOH 6mA VOL 0 4V IOL 6mA All Inputs are pulled high to 5 Vdc through a 10k resistor...

Page 23: ...p 11 Elite Series Pump 1 Connect the pump to the computer via USB Open the control panel and select System and Security and then System Click Device Manager on the left hand menu 2 Two Flash Update Applications will be displayed in the device manager under Other Devices Right click on either of the flash update applications and select Update Driver Software Pump 11 Elite OEM Module USER S MANUAL p...

Page 24: ...then click Next 5 If a warning message regarding the driver publisher is displayed choose Install this driver software anyway USB VIRTUAL COMMPORT DRIVER INSTALLATION CONTINUED Pump 11 Elite OEM Module USER S MANUAL phone 508 893 8999 fax 508 429 5732 e mail bioscience harvardapparatus com web www harvardapparatus com Pump 11 Elite OEM Module Series User s Manual Publication 5420 007 Rev B 6 Click...

Page 25: ...S 485 OUT PUMP ADDR 1 RS 485 OUT RS 485 IN PUMP ADDR n RS 485 IN RS 485 IEEE 1394 Cable RS 485 IEEE 1394 Cable When applying power to pumps in a daisy chain apply power to the last group and work backwards When removing power be sure to remember that power removed from a stand alone unit will also cause power to be removed from all RS 485 powered units deriving power from it Power may be removed f...

Page 26: ...g with Legacy RJ 11 Option Pump Transmit Port 2 out Receive Ground Ground Transmit Receive Port 1 in Legacy RJ 11 Option Pumps with Legacy RJ 11 Option P C PUMP ADDR Ø RJ 11 IN RJ 11 OUT PUMP ADDR 1 RJ 11 OUT RJ 11 IN PUMP ADDR n RJ 11 IN RS 232 D Sub 9 F to Telephone Cable Cat 70 2022 RS 232 Non Standard Telephone CableCat 55 7760 or 72 2478 RS 232 Non Standard Telephone Cable Cat 55 7760 or 72 2...

Page 27: ...hould be lubricated with Super Lube Synthetic Grease provided with the pump To clean the exterior surfaces use a lint free cloth to remove loose dust For more efficient cleaning use a soft cloth dampened not soaked with water an aqueous solution of 75 isopropyl alcohol or a mild detergent Pump 11 Elite OEM Module USER S MANUAL phone 508 893 8999 fax 508 429 5732 e mail bioscience harvardapparatus ...

Page 28: ...sly saved on your desktop and click Open 6 Click Start Update 7 After the update is complete press Reboot to power cycle the pump Pump 11 Elite OEM Module USER S MANUAL phone 508 893 8999 fax 508 429 5732 e mail bioscience harvardapparatus com web www harvardapparatus com Pump 11 Elite OEM Module Series User s Manual Publication 5420 007 Rev B Prior to upgrading users will need to install the Pump...

Page 29: ...ad of the syringe than the pump is capable of delivering To restart the pump remove the cause of stalling and or increase the force setting then press RUN The pump will continue the interrupted procedure where it stopped 3 Out of Range A value was entered or encountered in a pump Method that was beyond the pump s limits 4 Overload Protection The Pump 11 Elite OEM Module Series incorporates resetta...

Page 30: ... ft 70 4002 USB Cable for PC to Pump Communication 5 m 16 ft 70 4003 Cable Assy Daisy chain Legacy RS 232 RJ 11 0 5 m 2 ft 55 7760 Cable Assy Daisy chain Legacy RS 232 RJ 11 2 m 7 ft 72 2478 Line Cord US 115 v 5113 001 Adapter D sub 15 to Term Blk 70 4006 Footswitch w Phono Plug 70 2215 Adapter for 25ml 50ml Hamilton GasTight syringes 55 8000 Hex Key 3 32 5012 005 Lubricant SuperLube 1cc 70 4013 R...

Page 31: ...y Directive 2011 65 EU RoHS Directive Standard s to which conformity is declared Safety EN 61010 1 2012 3rd Edition Emissions Immunity EN 61326 2006 05 01 EN 61000 4 2 1995 A1 1998 A2 2001 EN 61000 4 3 2002 EN 61000 4 4 2004 EN 61000 4 5 1995 A1 2001 EN 61000 4 6 2007 EN 61000 4 8 1993 A1 2001 EN 61000 4 11 2004 EN 61000 3 2 2000 EN 61000 3 3 1995 A1 2001 CISPR11 2003 A1 2004 Group 1 Class A EMC a...