13
Operating the Pump 11 Elite Series
Pump 11 Elite OEM Module
USER’S MANUAL
phone
508.893.8999
fax
508.429.5732
web
www.harvardapparatus.com
Pump 11 Elite OEM Module Series User’s Manual
•
Publication 5420-007 Rev B
Error Messages
Error messages are displayed if the entered command
cannot be executed for some reason. The error message
will take up two lines with the first line being the
message type and the second line describing the error
itself. The second line may be up to 80 characters long.
Command Errors
Command errors are displayed when the command is
unrecognized, entered in the wrong mode, or the state
of the pump keeps the command from executing (i.e.
using the IRUN command if a limit switch is active).
The command error has the following format:
<lf>[##:]Command error:<cr>
<lf>[##:]<sp><sp><sp>{error message}<cr>
<lf>[##]<prompt>
Argument Errors
Argument errors are displayed when a command
argument is unrecognized or out of range. The argument
in question will be displayed except in the case of
missing arguments.
The argument error has the following format:
<lf>[##:]Argument error: [bad argument]<cr>
<lf>[##:]<sp><sp><sp>{error message}<cr>
<lf>[##]<prompt>
System commands
ADDRESS
Sets or displays the pump address. Valid range is 0 to
99.
Command format:
address [0-99]
Query response:
<lf>[##:]Pump address is #<cr>
<lf>[##]<prompt>
Notes: Pumps with an address of 0 are masters, and
pumps with an address between 1 and 99 are slaves.
BAUD
Sets or displays the baud rate. Valid baud rates are
9600, 19200, 38400, 57600, 115200, 128000, 230400,
256000, 460800, and 921600. Note that some
computers may not be able to handle baud rates above
115200.
Command format:
baud [9600 | 19200 | 38400 | 57600 | 115200 |
128000 | 230400 | 256000 | 460800 | 921600]
Query response:
<lf>[##:]# baud<cr>
<lf>[##]<prompt>
BOOT
Enter the bootloader.
Command format:
boot
CATALOG
Displays a catalog of Methods stored in the pump.
Command format:
cat
Query response:
<lf>[##:]Method name Size<cr>
<lf>[##:]--------------- ----<cr>
<lf>[##:]{Method name} #<cr> This line is
repeated for each file on the disk
<lf><lf>[##:]# file(s) using # steps<cr>
<lf>[##]<prompt>
Notes: The file size is the number of steps in the Method
plus one for the file information header.
The last line shows the number of files and the total
number of steps being used on the disk.
DELMETHOD
Deletes a Method stored in the pump.
Command format:
delMethod {Method}
ECHO
Sets or displays the USB echo state. Valid states are on
or off.
Command format:
echo [on|off]
Query response:
<lf>[##:] OFF<cr>
<lf>[##]<prompt>
or:
<lf>[##:] ON<cr>
<lf>[##]<prompt>
FORCE
Sets or displays the infusion force level in percent. Valid
range is 1 to 100.
Command format:
force [1-100]
Query response:
<lf>[##:]#%<cr>
<lf>[##]<prompt>