28
Appendices
Upgrading Pump 11 Elite Series Software
1. Upload the latest software version to your desktop (format is filename.srec).
2. Disconnect all I/O devices and then connect the pump to PC using a USB cable.
3. Using Method Manger, send the ‘Boot’ command to enter “Boot Loader Mode”.
4. From the CD provided with the pump, open the Pump 11 Elite Series Pump Updater Application.
5. Click Open Update File and browse to the file previously saved on your desktop and click Open.
6. Click Start Update.
7. After the update is complete, press Reboot to power cycle the pump.
Pump 11 Elite OEM Module
USER’S MANUAL
phone
508.893.8999
fax
508.429.5732
web
www.harvardapparatus.com
Pump 11 Elite OEM Module Series User’s Manual
•
Publication 5420-007 Rev B
Prior to upgrading, users will need to install
the Pump 11 Elite Series USB driver as well
as the bootloader driver file (Pump 11 Elite
Series-Bootloader.inf). Reference “USB
Virtual Comport Driver Installation”.
To exit boot loader mode without upgrading
the software, power cycle the pump.