8. Measuring in object mode
A distinction is made here between two different types of measurement. Follow
the steps below if you want to measure the surface temperature of an object,
e.g. a baby bottle or the surface temperature of the bathwater:
1. Please check that the mode switch is in the right position.
Follow
steps 1 – 2, as described in Chapter 7.
2. The temperature display then goes out. On the display, you will see the object
symbol
. The device is ready for the measurement as soon as the "°C"
symbol flashes.
3. Point the thermometer at the middle of the object you would like to measure
at a distance of 3 – 5 cm.
4. Press the SCAN button to start the measurement.
5. The scanning process lasts 3 seconds and the measured temperature then
appears on the blue illuminated display. The result is displayed for 5 seconds
and then the flashing "°C" symbol appears in the display. The device is now
ready for the next measurement.
If you want to measure the ambient or room temperature,
• hold the device in the room and press the SCAN button. You need not point the
device at any specific object. Thermoval baby now determines the temperature
of the ambient air. After 3 seconds, the temperature appears on the display
and the display lights up blue.
Important information for object measurement
• Due to physical surface effects, the temperature of a surface measured in
object mode can differ greatly from the real temperature.
• Inside the object, the real temperature may be distinctly warmer or colder than
the temperature measured on its surface.
• The measurement is accurate only when the measuring distance of 3 – 5 cm
between the thermometer and measured object is maintained and the sensor
lens is not clouded by condensation.
• Please make sure, before taking the measurement, that the thermometer
has been placed in the same room as the object to be measured for at least
30 minutes. In case of high air humidity (e.g. in the bathroom), the thermom-
eter must be acclimatised to the room temperature and air humidity prior to
measurement.
• Measurements above boiling water or steaming bathwater will cause the
sensor lens to become steamed up, and thus will have a strong influence on
the measuring accuracy.
• Surface measurement in object mode is not suitable for medical use or body
4_SpKr_Thermoval_baby_GA_925094_1.indd 78
18.08.15 13:31
Summary of Contents for 925094
Page 1: ...4_SpKr_Thermoval_baby_GA_925094_1 indd 109 18 08 15 13 31...
Page 4: ...1 5 4 6 7 3 2 4_SpKr_Thermoval_baby_GA_925094_1 indd 2 18 08 15 13 31...
Page 23: ...SI Sloven ina 21 4_SpKr_Thermoval_baby_GA_925094_1 indd 21 18 08 15 13 31...
Page 55: ...TR T rk e 53 4_SpKr_Thermoval_baby_GA_925094_1 indd 53 18 08 15 13 31...
Page 56: ...HK 54 4_SpKr_Thermoval_baby_GA_925094_1 indd 54 18 08 15 13 31...
Page 58: ...HK 56 4_SpKr_Thermoval_baby_GA_925094_1 indd 56 18 08 15 13 31...
Page 59: ...HK 57 4_SpKr_Thermoval_baby_GA_925094_1 indd 57 18 08 15 13 31...
Page 60: ...HK 58 4_SpKr_Thermoval_baby_GA_925094_1 indd 58 18 08 15 13 31...
Page 61: ...HK 59 4_SpKr_Thermoval_baby_GA_925094_1 indd 59 18 08 15 13 31...
Page 62: ...HK 60 4_SpKr_Thermoval_baby_GA_925094_1 indd 60 18 08 15 13 31...
Page 63: ...HK 61 4_SpKr_Thermoval_baby_GA_925094_1 indd 61 18 08 15 13 31...
Page 64: ...HK 62 4_SpKr_Thermoval_baby_GA_925094_1 indd 62 18 08 15 13 31...
Page 65: ...HK 63 4_SpKr_Thermoval_baby_GA_925094_1 indd 63 18 08 15 13 31...
Page 66: ...HK 64 4_SpKr_Thermoval_baby_GA_925094_1 indd 64 18 08 15 13 31...
Page 67: ...HK 65 4_SpKr_Thermoval_baby_GA_925094_1 indd 65 18 08 15 13 31...
Page 68: ...HK 66 4_SpKr_Thermoval_baby_GA_925094_1 indd 66 18 08 15 13 31...
Page 69: ...HK 67 4_SpKr_Thermoval_baby_GA_925094_1 indd 67 18 08 15 13 31...
Page 70: ...HK 68 4_SpKr_Thermoval_baby_GA_925094_1 indd 68 18 08 15 13 31...
Page 71: ...HK 69 4_SpKr_Thermoval_baby_GA_925094_1 indd 69 18 08 15 13 31...
Page 90: ...BG pc 88 1 HARTMANN Thermoval baby 4_SpKr_Thermoval_baby_GA_925094_1 indd 88 18 08 15 13 31...
Page 94: ...BG pc 92 17 10 5 Thermoval baby 6 4_SpKr_Thermoval_baby_GA_925094_1 indd 92 18 08 15 13 31...
Page 96: ...BG pc 94 30 10 8 1 4_SpKr_Thermoval_baby_GA_925094_1 indd 94 18 08 15 13 31...
Page 100: ...BG pc 98 14 2 15 3 HARTMANN 19 4_SpKr_Thermoval_baby_GA_925094_1 indd 98 18 08 15 13 31...
Page 105: ...BG pc 103 4_SpKr_Thermoval_baby_GA_925094_1 indd 103 18 08 15 13 31...