Chapter 4 - Setup
4-21
The least significant 16 bits are a command for Chan-
nel 0, the next 16 bits are a command for Channel 1.
Commands are “1 shot”, occurring upon a 0-1 transi-
tion.
Bit 0x2000 in the STATUSWORD will be set upon
completion of the output table comand. Bit 0x4000
will also be set if the command failed. Setting the
command word to zero will clear these status bits.
Error Code List
SUCCESS
0
OUTOFTOLERANCE
-3
NOSUCHCMD
-5
C2FAILNODEVS
-6
C2FAILCAPEQ
-7
(failure, capacities not
equal)
HARCALFAILCOUNTS -8
(failure, not enough
ADC counts between
high, low)
Timed Out
Commands
The ASIC on the HI 1756-WS module requires that a
command receive a reply within 65 milliseconds.
This is not enough time to complete some of the com-
mands. The commands subject to time-out are:
•
CALLOWCMD
•
C2CALCMD
•
WEIGHSYSTEST
•
C2SEARCH
The reply to these commands is sent before the com-
mands are actually performed.
To see if a calibration has succeeded, the command
“GETSTATUS” can be sent
The data returned by the WEIGHSYSTEST and
C2SEARCH commands is actually the data deter-
mined by an earlier command. In practice, these com-
mands need to be sent twice.
Return Codes
See Chapter 7, Troubleshooting.
Summary of Contents for A Series
Page 14: ...HI 1756 WS WEIGH SCALE MODULE 1 6 ...
Page 28: ...HI 1756 WS WEIGH SCALE MODULE 3 10 ...
Page 52: ...HI 1756 WS WEIGH SCALE MODULE 4 24 ...
Page 62: ...HI 1756 WS WEIGH SCALE MODULE 6 2 FIG 6 2 DISCRETE DATA ...
Page 72: ...HI 1756 WS MANUAL 7 10 ...
Page 80: ...HI 1756 WS MANUAL Write Parameters WRITEPARAM 4 18 Z Zero Command ZEROCMD 4 12 ...