E6000 Portable Multi-Gas Detector Henan Hanwei Electronics Co., Ltd.
6
Note:
If self-test fails, related information will be displayed. For
details, please refer to Clause 10 Troubleshooting of this
manual.
If self test succeeds, the device enters into warm-up period of
3-30s, which depends on the sensor type.
5.2 Power off
In normal detection status, hold
for 3 seconds and the
screen will show
“Shutting down…”. Meanwhile, the buzzer will give
out intermittent beep twice. Then machine is power off.
5.3 Bump test
Every day before using the device, the user is suggested carry out
Bump test, so as to check if the device is working normally.
Test method:
When the device is power on, put it into high level gas environment,
which is higher than the preset high alarm point of the device. If all
the device
’s function is ok, then device can be used in the working
area.
Note:
If any reading on the screen is beyond the prescribed display
error range, please follow Clause 5.7 and 5.8 of this manual to
re-calibrate it.
If detector does not response or showing errors, please contact
the seller for repairing.