Hanwei E6000 Operation Manual Download Page 10

E6000 Portable Multi-Gas Detector        Henan Hanwei Electronics Co., Ltd. 

 

Menu 

Submenu 

Remarks 

 

SAFETY 

DEPLOY

 

 

To set user

’s ID.   

 

ZERO 

CAL.

 

 

Zero calibration 

 

DATA   

UPLOAD

 

 

Data upload 

 

PERIPHE 

CONFIG. 

 

One mode must be selected 

 

POWER 

MANAGE

 

 

Activate the other function 

 

LANGUAGE 

SELECT 

 

One mode must be selected. 

Note: Initial password for entering advanced menu is 

“0000”. 

 

Summary of Contents for E6000

Page 1: ...E6000 Portable Multi Gas Detector Operation Manual Ver HW1403180296 Read this manual carefully before using this device...

Page 2: ...5 8 of this manual Periodic Bump test will test the response feature of the sensor Please make sure that the visual audible and vibrative alarm signals are ok Only accessories which are specified for...

Page 3: ...ry should be discarded by qualified withdrawers It s forbidden to disassembly adjust or repair the device without permission Please avoid the device falling from high place or serious shocking Any oth...

Page 4: ...s Design of self test self diagnose and self repairing Password management Intrinsically safe design 3 Technical specification Detecting method Natural diffusion Target gas Refer to the Annex 1 in the...

Page 5: ...lti Gas Detector Henan Hanwei Electronics Co Ltd 4 Explosion proof grade Exia IIC T4 Ga Ingress protection IP66 Dimensions and weight 168mm 91mm 45mm about 500g 4 Structure and functions Appearance Di...

Page 6: ...or more than 3s and the device will be power on The screen shows as below After power on the detector will proceed self test and display is as below drawing The device will make self test on the buzze...

Page 7: ...rmittent beep twice Then machine is power off 5 3 Bump test Every day before using the device the user is suggested carry out Bump test so as to check if the device is working normally Test method Whe...

Page 8: ...e enters into common menu setting interface Below is the button function Menu Submenu Remarks DATE TIME When changing the date or time the selected bit will blink ALARM MODE When changing the mode the...

Page 9: ...er can also exit without saving changes by pressing 5 4 2 Advanced menu In the interface of common menu interface press both and simultaneously twice The screen shows password inputting interface Inpu...

Page 10: ...bmenu Remarks SAFETY DEPLOY To set user s ID ZERO CAL Zero calibration DATA UPLOAD Data upload PERIPHE CONFIG One mode must be selected POWER MANAGE Activate the other function LANGUAGE SELECT One mod...

Page 11: ...zero drift of the gas sensors If finding the reading in clean air is not zero please make zero calibration by following Clause 6 7 and 6 8 of this manual 5 6 Status review In normal detection mode pre...

Page 12: ...5 8 1 Enter the setting interface While the device is power off hold both and buttons simutaneously for about 3 seconds Then the device first performs a self test as after power on and after a short...

Page 13: ...rns black Then press to enter the submenu 5 8 2 Zero calibration Select icon and press to enter auto zero calibration interface shown as below When finished a icon displays for gas succeed and for gas...

Page 14: ...alibration manually After that a icon displays for gas succeed and for gas failed shown as below 5 8 3 Span Calibration After zero calibration the device displays a countdown interface shown as below...

Page 15: ...d press to change the figure After changing the figure press button to save and enter the gas inputting mode shown as below When the device senses the input gas it displays as below During this period...

Page 16: ...manual calibration Once finishing calibration one sensor the device starts to calibrate the next sensor one by one The operation is same as above During calibration period press and the user can skip...

Page 17: ...move the cursor After adjusting press button to save Then the device enters setting interface of the next alarm point Note H High alarm point L Low alarm point ST STEL alert point TW TWA alert point 5...

Page 18: ...11 Password setting Move the cursor to the incon and press button to enter the submenu shown as below Press to adjust the figure and press button to move the cursor After setting press to save the ne...

Page 19: ...and USB data cable Connect the USB data calbe correctly between the device and computor Then run the matching program as insturcted in the manual for the software For details please refer to the softw...

Page 20: ...it immediately System breakdown Contact the seller Circuit fault Contact the seller No response to gas Warm up not finished Wait till it finishes Circuit fault Contact the seller Reading of gas level...

Page 21: ...se the device to other harsh environments including excessive cold heat humidity electromagnetc field and intense light Clean the housing of the device by using damp cloth Do not use corrosive agents...

Page 22: ...LEL 50 LEL N A N A H2 0 100 LEL 20 LEL 50 LEL N A N A H2 0 1000ppm 35ppm 250ppm N A N A O2 0 30 vol 19 5 vol 23 5 vol N A N A H2S 0 100ppm 10ppm 15ppm 10 15 CO 0 1000ppm 35ppm 200ppm 35 200 CO 0 2000...

Page 23: ...E6000 Portable Multi Gas Detector Henan Hanwei Electronics Co Ltd 22...

Page 24: ...HENAN HANWEI ELECTRONICS CO LTD No 169 Xuesong Road National Hi Tech Zone Zhengzhou 450001 Henan China Phone 86 371 67169070 67169080 E mail sales hwsensor com Website www hwsensor com...

Reviews: