E6000 Portable Multi-Gas Detector Henan Hanwei Electronics Co., Ltd.
1
SAFETY INFORMATION
Incorrect operation or unsuitable using environment may attenuate
the instrument’s performance. Therefore, please first read the
below safety information carefully before using and then operate
this device.
Please don't use the defective device. Before using, please
check if there is crack or spare part missing. If yes, please
contact the seller immediately.
It's suggested that the user carry out the "Bump Test" by
following Clause 5.3 of this manual before using the device. If
the device reading is beyond the specified range, please
calibrate the device by following Clause 5.7 and 5.8 of this
manual.
Periodic Bump test will test the response feature of the sensor.
Please make sure that the visual, audible and vibrative alarm
signals are ok.
Only accessories which are specified for E6000 or permitted by
the seller are allowed to be used.
Only the charger which is specified for E6000 is allowed to be
used. It's forbidden to charge the device in the dangerous
environment.
Please don't expose the device to the exceeding-range gas
environment for long time. Otherwise, it will badly influence the
performance and even damage the device.
If exposed to the environment consisting of leaded compound,
sulfocompound, organic phosphorus compound or silicon, the
gas sensor will be poisoned. Please don't use the device in the
above environment.
Please don't expose the device to the environment which
consists of H2S, hydrocarbons gas or high corrosive gas for