Media-Mate Plus Operation Manual
25-710-800-O
Test Cycle
Test Cycles are accomplished with the Dispense Head positioned on the Media Tank Cover
docking position. This arrangement provides for media return back to the Tank during an
instrument cycle. Test Cycles are useful for observation of instrument performance.
Running a Test Cycle:
1.
Set all user inputs (see
Section 7 – “Volume Selection,” “Temperature Setting,” and “Fill
Time Setting”
).
2.
Position the Dispense Head in the docking position on the Media Tank Cover.
3.
Fill the Media Tank with media.
4.
Plug the retractable power cord into an appropriate AC power receptacle (see
Section 5
– “Electrical”
for retracting requirements of cordage).
5.
Turn on Air Pump power at the Control Panel.
6.
Press the start button on the Control Panel.
Filling Dissolution Vessels
1.
Set all user inputs (see
Section 7 – “Volume Selection,” “Temperature Setting,” and “Fill
Time Setting”
).
2.
Fill the Media Tank with media.
3.
Plug the retractable power cord into an appropriate AC power receptacle (see
Section 5
– “Electrical”
for retracting requirements of cordage).
4.
Turn on the main power switch, located on the pump panel next to the power cord outlet.
5.
Turn on the Air Pump and Heater power switches, located on the Control Panel.
6.
Configure the Dispense Head (see
Section 7 – “Configuring the Dispense Head”
).
7.
Place the Dispense Head over the dissolution vessels on the Test Station.
8.
Verify that the active output ports on the Dispense Bar engage appropriate vessels on the
Dissolution Test Station.
9.
Verify the media temperature at the Control Panel.
10. Press the start button at the Control Panel.
11. When the cycle is finished, remove the Dispense Head from the Dissolution Test Station
and return to the docking position on the Media Tank Cover.
Section 7
Operating Instructions
29