Hanna FC 300 Instruction Manual Download Page 6

10

11

XII. 

XII. 

XII. 

XII. 

XII.  Other Measurement T

Other Measurement T

Other Measurement T

Other Measurement T

Other Measurement Techniques

echniques

echniques

echniques

echniques

Known Addition  (for Na

+

)

An unknown concentration can be determined by adding a
known amount (volume and concentration) of measured
ion to a known volume of the sample. This technique is
called Known Addition.   The method can use an ideal
sensor slope, but actual determined slopes at the tempera-
ture of measurement should be used if known.  This method
is preprogrammed in the Hanna HI 4222 pH/ISE/mV meter,
which simplifies the method greatly.
Example: Sodium ion determination with known addition.
1. A 50 mL sample of unknown (Vsample) is placed in

a clean plastic beaker with a FC300 sensor. Add 10
mL of ISA HI 4016-00 or  HI7090L (V

ISA)

   Mix well.

and record the stable mV value. (mV 1)

2. 5 mL (Vstd) of 10

-1

M (Cstd) standard is added to the

beaker and the mV value increases.  The unknown
sodium concentration in the original sample
(Csample) can then be determined by the following
equation.

3. The procedure can be repeated with a second stan-

dard addition to verify slope and operation of the
method.

XIII.

XIII.

XIII.

XIII.

XIII.pH

pH

pH

pH

pH

The FC300 electrodes are best used in pH from 9.5 to 14.00.
The ratio of sodium to hydrogen ion should be greater than
10000. For example, if  sodium is .05M, Hydrogen ion
must be 0.000005M; 5.30 pH or greater. If sodium is
.00001M, pH must be  7.00 pH. Samples that fall beyond
this range must be adjusted with ISA. See Section VI.

XIV

XIV

XIV

XIV

XIV. Storage

Storage

Storage

Storage

Storage

The FC300 sodium electrodes may be stored in storage
solution or in aqueous sodium solutions with ISA between
sample measurements.
If the electrode will be used frequently and needs to be
ready for use, take measures to prevent evaporation of fill
solution. Top off fill solution, and replace the fill hole cover
and  protective cap with a few drops of storage solution over
the sodium sensitive membrane . Store electrode upright.
Before using, rinse off electrode and top off fill solution.
For long term storage,  wash all salts from assembly with
deionized water. Replace the fill hole cover and  protective
cap with a few drops of storage solution over the sodium
sensitive membrane. Store in shipment box.

 C

sample 

=

(

V

T

)10

E/S 

- (

V

S

)

C

standard

V

standard

V

sample

V

S

(

V

sample

+

V

standard

+

V

ISA

)= 

V

T

(

V

sample

+

V

ISA

)= 

V

S

XV

XV

XV

XV

XV.  

.  

.  

.  

.  Conversion T

Conversion T

Conversion T

Conversion T

Conversion Tables

ables

ables

ables

ables

          Multiply by

          Multiply by

          Multiply by

          Multiply by

          Multiply by

Moles/L (M) Na

+

 to ppm Na

+

 (mg/L)

22990

ppm Na

+

 (mg/L) to M Na

+

(Moles/L)

4.35 X 10

-5

-5

-5

-5

-5

ppm Na

+

 (mg/L) to ppm NaCl (mg/L)

2.54197

ppm NaCl (mg/L) to ppm Na

+

 (mg/L)

0.39339

Summary of Contents for FC 300

Page 1: ...1 FC 300 FC 300 FC 300 FC 300 FC 300 Combination Sodium Electrode Instruction Manual ...

Page 2: ... The FC300 sodium electrode is a potentiometric device used for the rapid determination of free sodium ions in water soft drinks beer wine and soils The electrode functions as a sensor or ionic conductor The FC300 is a combination electrode that does not require a separate reference electrode to complete its electrolytic circuit The selective glass membrane blown on the tip of the sensor exchanges...

Page 3: ...ed Solutions Required Solutions Required Standards for Sodium Measurements Select appropriate Hanna Instruments standard and ISA from the list below Part number 0 1 M sodium solution HI4016 01 10 ppm sodium solution HI4016 10 100 ppm sodium solution HI4016 02 1000 ppm sodium solution HI4016 03 30 0g L NaCl solution 500mL HI 7081L 3 0g L NaCl solution 500mL HI 7083L 0 3g L NaCl solution 500mL HI 70...

Page 4: ...one minute Discard conditioning solution after use do not return to bottle Rinse off sensor bulb in water and place into a smallcontainerofstoragesolution HI4016 45 This so lution provides optimum conditions to rehydrate the mem brane surface and prepare it for sodium measurements A minimum of one hour is recommended for rehydrating the sensing surface Calibration standards and sample solutions sh...

Page 5: ...um specialprecautionsmustbeemployedfor reproduciblemeasurements Waterusedforstandardsmust be sodium free and sensors and glassware must be rinsed repeatedly with this water to prevent carry over In the region where the electrode response appears curved more calibration points are needed and calibration will need to be repeated more frequently A pH mV meter in mV mode with semi log graph paper may ...

Page 6: ... The ratio of sodium to hydrogen ion should be greater than 10000 For example if sodium is 05M Hydrogen ion must be 0 000005M 5 30 pH or greater If sodium is 00001M pH must be 7 00 pH Samples that fall beyond this range must be adjusted with ISA See Section VI XIV XIV XIV XIV XIV Storage Storage Storage Storage Storage The FC300 sodium electrodes may be stored in storage solution or in aqueous sod...

Page 7: ...teofpurchase when used for their intended purpose and maintained according to instructions If they fail to work when first used contact your dealer immediately Damage due to accidents misuse misapplication tampering or lack of prescribed maintenance is not covered HannaInstrumentsreservestherighttomodifythedesign construction or appearance of its products without advance notice ...

Reviews: