Hanna FC 300 Instruction Manual Download Page 4

6

7

VII.

VII.

VII.

VII.

VII. General Guidelines

General Guidelines

General Guidelines

General Guidelines

General Guidelines

During measurement always operate electrode with
the fill hole open.

Verify protective cap has been removed.

During normal use, fill solution will slowly drain out
of the ceramic junction located on the side of the  the
electrode. Excessive loss (>2 cm drop within 24 hours)
is not normal.

Add fill solution daily to maintain a good head
pressure. For optimum reference response, this level
should be maintained and not be allowed to drop
more than 2-3 cm (1-inch) below fill hole.

VIII.

VIII.

VIII.

VIII.

VIII. Electrode Preparation

Electrode Preparation

Electrode Preparation

Electrode Preparation

Electrode Preparation

 Before using the electrode  for the first time or if reactivating
it after  storage,  the electrode should be conditioned, soaked
in storage  solution and then tested.
For a new sensor:
Remove the protective shipping cap from the glass bulb
and save for storage.
Slide fill hole cover off opening to refill reference cavity
with HI 7079 fill solution. Leave open during measure-
ment.
Condition/etch sodium sensitive membrane:

Warning: Use extreme care when handling condtioning
solution. Protective  gloves and eyeware should be worn.

This procedure removes a thin layer of sensitive glass from
the sodium membrane exposing a fresh surface. Pour a
small quantity of Hanna HI 4016-46 conditioning solution
into a plastic container. Place the sensing membrane (round
bulb at tip of sensor) into the solution for one minute.
Discard conditioning solution after use , do not return to
bottle.  Rinse off sensor bulb in water and place into a
small container of storage solution (HI 4016-45). This  so-
lution provides optimum conditions to rehydrate the mem-
brane  surface and prepare it for sodium measurements. A
minimum of one hour is recommended for rehydrating the
sensing surface.

Calibration standards and sample solutions should
have the same ionic strength. ISA should be added to
both samples and standards immediately before tak-
ing measurements.

•    

Calibration standards and sample solutions should be

at the same temperature. Thermally insulate solution
vessel from magnetic stirrer with cork or other insulat-
ing medium.

Calibration standards and sample solutions should
be stirred at the same rate using identical sized stir
bars.

Surface coating  will effect the response.  Inspect sen-
sor before using.

Rinse electrode with distilled or deionized water be-
tween samples and dab dry with lab wipe or other soft
disposable absorbent toweling.

Check calibration every 1-2 hours.

IX.   

IX.   

IX.   

IX.   

IX.   Quick Check of Electrode Slope

Quick Check of Electrode Slope

Quick Check of Electrode Slope

Quick Check of Electrode Slope

Quick Check of Electrode Slope

Connect sensors to pH/mV/ISE meter

Place meter in mV mode.

Place 100 mL of DIW into a beaker with 10 mL ISA
and a stir bar.

Place electrode into prepared sample.

Add 1 mL of a standard (0.1 M or 1000 ppm stan-
dard) to beaker. Record the mV value when stable.

Add an additional 10-mL of standard to the solution.
Record the mV when reading has stabilized.  This
value should be more positive than the previous noted

Determine the difference between the two-mV values.
An acceptable value for this slope is 56 ± 4 mV.

Summary of Contents for FC 300

Page 1: ...1 FC 300 FC 300 FC 300 FC 300 FC 300 Combination Sodium Electrode Instruction Manual ...

Page 2: ... The FC300 sodium electrode is a potentiometric device used for the rapid determination of free sodium ions in water soft drinks beer wine and soils The electrode functions as a sensor or ionic conductor The FC300 is a combination electrode that does not require a separate reference electrode to complete its electrolytic circuit The selective glass membrane blown on the tip of the sensor exchanges...

Page 3: ...ed Solutions Required Solutions Required Standards for Sodium Measurements Select appropriate Hanna Instruments standard and ISA from the list below Part number 0 1 M sodium solution HI4016 01 10 ppm sodium solution HI4016 10 100 ppm sodium solution HI4016 02 1000 ppm sodium solution HI4016 03 30 0g L NaCl solution 500mL HI 7081L 3 0g L NaCl solution 500mL HI 7083L 0 3g L NaCl solution 500mL HI 70...

Page 4: ...one minute Discard conditioning solution after use do not return to bottle Rinse off sensor bulb in water and place into a smallcontainerofstoragesolution HI4016 45 This so lution provides optimum conditions to rehydrate the mem brane surface and prepare it for sodium measurements A minimum of one hour is recommended for rehydrating the sensing surface Calibration standards and sample solutions sh...

Page 5: ...um specialprecautionsmustbeemployedfor reproduciblemeasurements Waterusedforstandardsmust be sodium free and sensors and glassware must be rinsed repeatedly with this water to prevent carry over In the region where the electrode response appears curved more calibration points are needed and calibration will need to be repeated more frequently A pH mV meter in mV mode with semi log graph paper may ...

Page 6: ... The ratio of sodium to hydrogen ion should be greater than 10000 For example if sodium is 05M Hydrogen ion must be 0 000005M 5 30 pH or greater If sodium is 00001M pH must be 7 00 pH Samples that fall beyond this range must be adjusted with ISA See Section VI XIV XIV XIV XIV XIV Storage Storage Storage Storage Storage The FC300 sodium electrodes may be stored in storage solution or in aqueous sod...

Page 7: ...teofpurchase when used for their intended purpose and maintained according to instructions If they fail to work when first used contact your dealer immediately Damage due to accidents misuse misapplication tampering or lack of prescribed maintenance is not covered HannaInstrumentsreservestherighttomodifythedesign construction or appearance of its products without advance notice ...

Reviews: