The Hanna FC300  Sodium electrode is a glass combina-
tion ion selective electrode designed for the measurement
of sodium ions in aqueous solutions such as  water, food,
beverages, wine, beer and soil.
The electrode utilizes a special glass membrane that is
selective to sodium ions. The internal reference electrolyte
chamber is refillable.

II

IIII

IIII. Specifications

Specifications

Specifications

Specifications

Specifications

Type:

Glass combination ISE with

Ag/AgCl reference

Ion measured:

Sodium (Na

+

)

Measurement range:

1M to 1X 10

-5

M

39100 to 0.039ppm

Operating Temperature:

0-80

°

C

Operating pH:

9.75-14 pH

Dimensions:

12 mm (OD) x 120 mm
insertion (0.47

 x 4.72

)

Connection:

BNC Plug

Wetted Materials:

glass, ceramic

III.

III.

III.

III.

III. Theory of Operation:

Theory of Operation:

Theory of Operation:

Theory of Operation:

Theory of Operation:

The FC300 sodium electrode is a  potentiometric device
used for the rapid determination of free sodium ions in
water, soft drinks, beer, wine, and soils.  The electrode
functions as a sensor or ionic conductor.  The FC300 is a
combination electrode that does not require a separate
reference electrode to complete its electrolytic circuit.  The
selective glass membrane blown on the tip of the sensor
exchanges ions with the sample solution which  produces a
voltage on the surface of the membrane. This creates a
charge imbalance between the test solution and internal
cell of the sensor. This exchange on the external surface
produces a voltage that changes in response to the sample

s

ion activity. When the ionic strength of the sample is fixed,
and hydrogen ions are eliminated, the voltage is propor-
tional to the concentration of sodium ions in solution. The
sensor follows the Nernst Equation:

E = E

a   

+ 2.3 RT/nF log A 

ion

E = observed potential

E

a

 = Reference and fixed internal voltages

R = gas constant (8.314 J/K Mol)

n= Charge on ion (equivalents/mol) (1+)

ion 

= ion activity in sample

T = absolute temperature in

 

K

F = Faraday constant (9.648 x 10

C/equivalent)

Interference:
Ratio of interfering ion to Na

+

 must be below

.0001 for H 

+

1 for K 

+

 Note: other monovalent
cations may also
increase sodium
readings.

Summary of Contents for FC 300

Page 1: ...1 FC 300 FC 300 FC 300 FC 300 FC 300 Combination Sodium Electrode Instruction Manual ...

Page 2: ... The FC300 sodium electrode is a potentiometric device used for the rapid determination of free sodium ions in water soft drinks beer wine and soils The electrode functions as a sensor or ionic conductor The FC300 is a combination electrode that does not require a separate reference electrode to complete its electrolytic circuit The selective glass membrane blown on the tip of the sensor exchanges...

Page 3: ...ed Solutions Required Solutions Required Standards for Sodium Measurements Select appropriate Hanna Instruments standard and ISA from the list below Part number 0 1 M sodium solution HI4016 01 10 ppm sodium solution HI4016 10 100 ppm sodium solution HI4016 02 1000 ppm sodium solution HI4016 03 30 0g L NaCl solution 500mL HI 7081L 3 0g L NaCl solution 500mL HI 7083L 0 3g L NaCl solution 500mL HI 70...

Page 4: ...one minute Discard conditioning solution after use do not return to bottle Rinse off sensor bulb in water and place into a smallcontainerofstoragesolution HI4016 45 This so lution provides optimum conditions to rehydrate the mem brane surface and prepare it for sodium measurements A minimum of one hour is recommended for rehydrating the sensing surface Calibration standards and sample solutions sh...

Page 5: ...um specialprecautionsmustbeemployedfor reproduciblemeasurements Waterusedforstandardsmust be sodium free and sensors and glassware must be rinsed repeatedly with this water to prevent carry over In the region where the electrode response appears curved more calibration points are needed and calibration will need to be repeated more frequently A pH mV meter in mV mode with semi log graph paper may ...

Page 6: ... The ratio of sodium to hydrogen ion should be greater than 10000 For example if sodium is 05M Hydrogen ion must be 0 000005M 5 30 pH or greater If sodium is 00001M pH must be 7 00 pH Samples that fall beyond this range must be adjusted with ISA See Section VI XIV XIV XIV XIV XIV Storage Storage Storage Storage Storage The FC300 sodium electrodes may be stored in storage solution or in aqueous sod...

Page 7: ...teofpurchase when used for their intended purpose and maintained according to instructions If they fail to work when first used contact your dealer immediately Damage due to accidents misuse misapplication tampering or lack of prescribed maintenance is not covered HannaInstrumentsreservestherighttomodifythedesign construction or appearance of its products without advance notice ...

Reviews: