ACCESSORIES

pH CALIBRATION SOLUTIONS
HI 50003-01

pH 3.00 buffer solution, 20 mL sachet, 10 pcs

HI 50003-02

pH 3.00 buffer solution, 20 mL sachet, 25 pcs

HI 70007P

pH 7.01 buffer solution, 20 mL sachet, 25 pcs

HI 5003

pH 3.00 buffer solution, 500 mL bottle

HI 7006L

pH 6.86 buffer solution, 500 mL bottle

HI 7007L

pH 7.01 buffer solution, 500 mL bottle

HI 8006L

pH 6.86 buffer solution in FDA bottle, 500 mL

HI 8007L

pH 7.01 buffer solution in FDA bottle, 500 mL

ELECTRODE STORAGE SOLUTION
HI 70300L

Storage solution, 500 mL bottle

HI 80300L

Storage solution in FDA bottle, 500 mL

ELECTRODE CLEANING SOLUTIONS
HI 70000P

Electrode rinsing solution, 20 mL sachet, 25 pcs

HI 7061L

General cleaning solution, 500 mL bottle

HI 70635L

Cleaning solution for wine deposits, 500 mL bottle

HI 70636L

Cleaning solution for wine stains, 500 mL bottle

HI 7073L

Protein cleaning solution, 500 mL bottle

HI 7074L

Inorganic cleaning solution, 500 mL bottle

HI 7077L

Oil & Fat cleaning solution, 500 mL bottle

HI 8061L

General cleaning solution in FDA bottle, 500 mL

HI 8073L

Protein cleaning solution in FDA bottle, 500 mL

HI 8077L

Oil & Fat Cleaning solution in FDA bottle, 500 mL

ELECTROLYTE SOLUTION
HI 7082

3.5M KCl electrolyte solution, 4 x 30 mL, for double junction
electrodes

HI 8082

3.5M KCl electrolyte solution in FDA bottle, 4 x 30 mL, for
double junction electrodes

ORP PRETREATMENT SOLUTIONS
HI 7091L

Reducing pretreatment solution, 500 mL bottle

HI 7092L

Oxidizing pretreatment solution, 500 mL bottle

Cleaning procedure

After use, clean the electrode by immersing it in the proper cleaning solution (

HI

700635

 or 

HI 700636

) for 5 minutes.

If case of slow and unreliable readings, it is recommended to perform the
following procedure for cleaning the junction:
1. Shift upwards the PTFE collar. If the collar does not shift easily, immerse it

into warm water.

2. Wipe gently the electrode surface with a soft and clean tissue.
3. In case of persistent wine or must residual, wet the tissue with cleaning

solution (

HI 700635

 or 

HI 700636

).

4. Put back the collar in its original position.

Troubleshooting

Evaluate your electrode performance based on the following.

Noise 

(Readings fluctuate up and down) could be due to:

– Clogged/Dirty Junction: 

refer to the cleaning procedure above.

– Loss of shielding

 due to low electrolyte level (in refillable electrodes

only): refill with fresh 

HI 7082 

solution.

Dry Membrane/Junction:

 soak in 

HI 70300

 storage solution

 

for 1 hour.

Drifting:

 soak the electrode tip in warm (approx. 50-60°C) 

HI 7082

solution for one hour and rinse the tip with distilled water. Refill with fresh

HI 7082 

solution.

Low Slope:

 refer to the cleaning procedure above.

No Slope: 

check the electrode for cracks in glass stem or bulb and replace

the electrode.

Slow Response/Excessive Drift: 

soak the tip in 

HI 7061 

solution for 30

minutes, rinse thoroughly in distilled water and then follow the cleaning
procedure above.

Summary of Contents for pH 209V

Page 1: ...Instruction Manual w w w h a n n a i n s t c o m pH 209V Complete Laboratory Kit for pH ORP Measurements in Wine Must and Grape Juice ...

Page 2: ...des pH 209 bench pH meter HI 1048B pH electrode HI 98501 electronic thermometer pH 3 and pH 7 buffer solutions 20 mL sachet each HI 700635 and HI 700636 specific cleaning solutions 20 mL each HI 7082 electrode refilling solution 30 mL bottle and syringe HI 76404 electrode holder 12 Vdc adapter and instruction manual Note Save all packing material until you are sure that the instrument functions co...

Page 3: ...le with BNC and 1 m 3 3 cable Input Impedance 1012 ohm Power Supply 12 Vdc adapter included Environment 0 to 50 C 32 to 122 F RH max 95 non condensing Dimensions 240 x 182 x 74 mm 9 4x7 2x2 9 Weight 1 0 kg 2 2 lb meter only SPECIFICATIONS 1 Liquid Crystal Display LCD 2 pH slope adjustment knob 3 pH offset adjustment knob 4 pH range selection key 5 mV range selection key 6 C range selection key 7 T...

Page 4: ...Vdc adapter into the power supply socket 9 on page 4 Note Make sure the main line is protected by a fuse Electrode connection For pH or ORP electrodes with internal reference connect the electrode to the BNC socket on the rear panel 10 on page 4 For electrodes with separate reference plug the BNC connector of the electrode to the BNC socket 10 on page 4 and the reference electrode to the proper so...

Page 5: ...rence Switch the instrument on by pressing the ON OFF button Rinse and immerse the pH electrode in pH 7 01 buffer and shake briefly Wait a couple of minutes for the reading to stabilize Note the temperature of the buffer solution using the supplied HI 98501 thermometer e g 20 C Press the C key to select temperature setting then adjust the Temperature knob until the LCD shows the noted temperature ...

Page 6: ...e the electrode Refilling the electrode ALWAYS VERIFY THE ELECTROLYTE LEVEL If the electrolyte level is more than 2 cm below the refilling hole unscrew and remove the refill ing hole cap Add the 3 5M KCl electrolyte solution HI 7082 by using the supplied syringe Tighten back the cap Wait at least 1 hour before using the electrode pH VALUES AT VARIOUS TEMPERATURES pH readings are strongly affected ...

Page 7: ...trodes ORP PRETREATMENT SOLUTIONS HI 7091L Reducing pretreatment solution 500 mL bottle HI 7092L Oxidizing pretreatment solution 500 mL bottle Cleaning procedure After use clean the electrode by immersing it in the proper cleaning solution HI 700635 or HI 700636 for 5 minutes If case of slow and unreliable readings it is recommended to perform the following procedure for cleaning the junction 1 Sh...

Page 8: ...g calibration of instruments ESD wrist straps should be worn to avoid possible damage to the electrode by electrostatic discharge To maintain the EMC performance of this equipment the recommended cables noted in the user s manual must be used Any variation introduced by the user to the supplied equipment may degrade the instrument s EMC performance To avoid electrical shock do not use this instrum...

Page 9: ...2942 Japan Tel 03 3258 9565 Fax 03 3258 9567 Korea Tel 02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 SouthAfrica Tel 011 615 6076 Fax 011 615 8582 Taiwan Tel 886 2 2739 3014 Fax 886 2 2739 2983 Thailand Tel 66 2619 0708 Fax 66 2619 0061 UnitedKingdom Tel 01525 850 855 Fax 01525 853 668 USA Tel 401 765 7500 Fax 401 765 7575 For e ma...

Reviews: