Hanna Instruments pH 209V Instruction Manual Download Page 4

7

6

4 cm

(1½")

• Press the pH key to display the pH measurement.
• The display shows the pH value of the test solution compensated for

temperature.

If measurements are taken in different samples successively, it is recommended
to rinse thoroughly the electrode with water, and then for some of the next
solution to be measured.

TAKING ORP MEASUREMENTS

pH 209 

can also measure ORP, using an ORP electrode. Hanna Instruments

offers a variety of ORP electrodes for this purpose. Contact your dealer for more
information.

• Connect the ORP electrode to the meter and sub-

merge the tip (4 cm / 1½") into the sample to be
tested.

Note:

ORP measurements are not compensated for temperature variation.

• Press the mV key to enter the mV mode and allow a few minutes for the reading

to stabilize.

• The display will show the ORP (mV) value.

AFTER MEASUREMENTS

• Press the ON/OFF button to switch the instrument off.

Power connection

Plug the 12 Vdc adapter into the power supply socket (#9 on page 4).

Note:

 Make sure the main line is protected by a fuse.

Electrode connection

For 

pH

 or 

ORP

 electrodes with internal reference, connect the electrode to the

BNC socket on the rear panel (#10 on page 4).
For electrodes with separate reference, plug the BNC connector of the electrode
to the BNC socket (#10 on page 4), and the reference electrode to the proper
socket (#11 on page 4) on the rear panel.

Note:

 To prevent damage to the electrode, remove the pH electrode from the

solution before turning the meter off. With the meter OFF, detach the
electrode and immerse it in the storage solution.

TAKING  pH MEASUREMENTS

Make sure that the instrument has been calibrated for pH before taking pH
measurements.

• Switch the instrument on by pressing the ON/OFF

button.

• Immerse the electrode tip (4 cm/1½") into the

sample and shake briefly.

• Take the temperature of the solution with the sup-

plied 

HI 98501

 thermometer (e.g. 25°C).

• Press the °C key to display temperature setting on the LCD and adjust the

temperature knob to display the temperature of the sample.

4 cm

(1½")

OPERATIONAL GUIDE

Summary of Contents for pH 209V

Page 1: ...Instruction Manual w w w h a n n a i n s t c o m pH 209V Complete Laboratory Kit for pH ORP Measurements in Wine Must and Grape Juice ...

Page 2: ...des pH 209 bench pH meter HI 1048B pH electrode HI 98501 electronic thermometer pH 3 and pH 7 buffer solutions 20 mL sachet each HI 700635 and HI 700636 specific cleaning solutions 20 mL each HI 7082 electrode refilling solution 30 mL bottle and syringe HI 76404 electrode holder 12 Vdc adapter and instruction manual Note Save all packing material until you are sure that the instrument functions co...

Page 3: ...le with BNC and 1 m 3 3 cable Input Impedance 1012 ohm Power Supply 12 Vdc adapter included Environment 0 to 50 C 32 to 122 F RH max 95 non condensing Dimensions 240 x 182 x 74 mm 9 4x7 2x2 9 Weight 1 0 kg 2 2 lb meter only SPECIFICATIONS 1 Liquid Crystal Display LCD 2 pH slope adjustment knob 3 pH offset adjustment knob 4 pH range selection key 5 mV range selection key 6 C range selection key 7 T...

Page 4: ...Vdc adapter into the power supply socket 9 on page 4 Note Make sure the main line is protected by a fuse Electrode connection For pH or ORP electrodes with internal reference connect the electrode to the BNC socket on the rear panel 10 on page 4 For electrodes with separate reference plug the BNC connector of the electrode to the BNC socket 10 on page 4 and the reference electrode to the proper so...

Page 5: ...rence Switch the instrument on by pressing the ON OFF button Rinse and immerse the pH electrode in pH 7 01 buffer and shake briefly Wait a couple of minutes for the reading to stabilize Note the temperature of the buffer solution using the supplied HI 98501 thermometer e g 20 C Press the C key to select temperature setting then adjust the Temperature knob until the LCD shows the noted temperature ...

Page 6: ...e the electrode Refilling the electrode ALWAYS VERIFY THE ELECTROLYTE LEVEL If the electrolyte level is more than 2 cm below the refilling hole unscrew and remove the refill ing hole cap Add the 3 5M KCl electrolyte solution HI 7082 by using the supplied syringe Tighten back the cap Wait at least 1 hour before using the electrode pH VALUES AT VARIOUS TEMPERATURES pH readings are strongly affected ...

Page 7: ...trodes ORP PRETREATMENT SOLUTIONS HI 7091L Reducing pretreatment solution 500 mL bottle HI 7092L Oxidizing pretreatment solution 500 mL bottle Cleaning procedure After use clean the electrode by immersing it in the proper cleaning solution HI 700635 or HI 700636 for 5 minutes If case of slow and unreliable readings it is recommended to perform the following procedure for cleaning the junction 1 Sh...

Page 8: ...g calibration of instruments ESD wrist straps should be worn to avoid possible damage to the electrode by electrostatic discharge To maintain the EMC performance of this equipment the recommended cables noted in the user s manual must be used Any variation introduced by the user to the supplied equipment may degrade the instrument s EMC performance To avoid electrical shock do not use this instrum...

Page 9: ...2942 Japan Tel 03 3258 9565 Fax 03 3258 9567 Korea Tel 02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 SouthAfrica Tel 011 615 6076 Fax 011 615 8582 Taiwan Tel 886 2 2739 3014 Fax 886 2 2739 2983 Thailand Tel 66 2619 0708 Fax 66 2619 0061 UnitedKingdom Tel 01525 850 855 Fax 01525 853 668 USA Tel 401 765 7500 Fax 401 765 7575 For e ma...

Reviews: