Separate reagent caps are provided in the accessory kit. Put
the supplied caps onto each reagent bottle prior to installing
them. Place the indicator bottle (HI 70498A) on the right and
the buffer bottle (HI 70498B) on the left.
Note:
Add the content of 5 HI 70452 sachets, DPD Compound, to the
Indicator Solution prior to installing it.
Connect the longer tube
ends on the left side of the
pump to the reagent bottle
cap insert fitting.
Connect the short ends in
the right side of the pump
to the measuring cell re-
agent input port through the
“Y” connector.
ELECTRICAL CONNECTIONS
A power cable (3 mt.) is provided with your analyzer. However,
if access to the terminal block is required, see below.
Warning:
Electrical connections should be installed only by qualified
personnel to assure conformity to applicable electrical codes.
Unplug the meter before any electrical connection.