8. WARNING AND ERROR MESSAGES
8.1. WARNING MESSAGES
FACTORY METHODS FULL
The maximum number of factory methods have been added.
USER METHODS FULL
The maximum number of user methods have been added. At least 1 user method
needs to be deleted before a new one can be created.
FAVORITE METHODS FULL
The maximum number of favorite methods have been added.
METHOD MISSING OR CORRUPT
Method file corrupted.
FILE MISSING OR CORRUPT
Log file corrupted.
DISK FULL FACTORY
Factory partition full.
DISK FULL
The maximum number of logs have been saved. At least 1 log needs to be deleted
before a new one can be created.
FLASH NOT SUPPORTED
USB flash drive not supported.
FLASH REMOVED
The USB flash drive is missing or cannot be read.
LOG CORRUPTED
Log file corrupted.
NO LIGHT
The Light Source is not functioning properly. Replace the lamp or check the wiring.
LOW LIGHT
The instrument cannot adjust the light level. Please check that the sample does not
contain any debris.
LIGHT HIGH
There is too much light to perform a measurement. Please check the preparation of
the zero cuvette.
REFERENCE ERROR
There is a problem with the reference chanel.
CLOSE THE LID
The lid is not properly closed.
INVERTED CUVETTES
The sample and the zero cuvettes are were measured in the wrong order, or there
is a problem with the cuvette preparation.
WRONG OR MISSING CUVETTE
Wrong cuvette inserted. The cuvette does not match the one specified in the
method.
NOT CALIBRATED
A calibration is required before a user method can be used.
INVALID CALIBRATION
The calculated slope for the calibration curve is outside of allowed range. Please
repeat the calibration.
HIGH TEMPERATURE
The internal temperature is higher then 55 °C.
LOW TEMPERATURE
The internal temperature is lower then 0 °C.
LAMP OLD - REPLACE SOON
The lamp life is over recommended maximum period. Consider replacing the lamp.
W
ARNING AND ERROR MESSAGES