5. The meter is ready for use. Always utilize the selected adapter for both "Zero" and "Read" measurements as specified in the
method instructions.
WARNING
: Improper use of the vial adapters could cause irreversible damage to the meter. Always use the flowing
precautions when using vial adapters.
•Never use excessive force to insert the adapter. You should be able to insert the vial with light pressure. If the vial
is not reaching the bottom, if there is large resistance, or if you are receiving a “light low” error during the “Zero”
operation, re-check that the indexing marks are aligned on the adapter and meter.
•Never insert hot vials/samples into the vial adapter. Samples should be near room temperature before inserting
into the meter/adapter.
•Do not attempt to close the sample cover while using the 13 mm vials or adapter. It is normal for the vials/
adapter to prevent the cover from closing completely.