background image

MAIN FEATURES

•  Simultaneous pH and temperature measurements on a 

large dual‑line LCD display

•  Automatic pH calibration at one or two points within 

two memorized buffer sets (standard or NIST)

•  Selectable temperature unit (°C or °F)
•  Electrode condition indicator
•  mV of pH measurement for electrode check
• 

FC10483

 dedicated pH with integrated temperature 

sensor

•  Probe quick connect system
•  Battery life indication and low battery detection
•  Keystroke confirmation tone
•  Auto‑off function
•  Waterproof casing IP67

MAIN FEA

TURES

Summary of Contents for HI99111

Page 1: ...HI99111 Portable Waterproof pH Temperature Meter for Wine Analysis INSTRUCTION MANUAL ...

Page 2: ... this meter as well as a precise idea of its versatility If you need additional technical information do not hesitate to e mail us at tech hannainst com or view our worldwide contact list at www hannainst com Dear Customer All rights are reserved Reproduction in whole or in part is prohibited without the written consent of the copyright owner Hanna Instruments Inc Woonsocket Rhode Island 02895 USA...

Page 3: ...nd Intended use Main features Specifications Display description Operational guide Meter setup Battery replacement Accessories Electrode maintenance Certification Recommendations for users Warranty TABLE OF CONTENTS 4 6 5 7 8 19 17 8 14 16 18 15 19 ...

Page 4: ... temperature probe with glass body and DIN connector with 1 m 3 3 cable HI50003 pH 3 00 buffer 1 sachet HI70007 pH 7 01 buffer 1 sachet HI700635 Cleaning solution for wine deposits 1 sachet HI700636 Cleaning solution for wine stains 1 sachet 100 mL beaker 1 pcs 1 5V AAA alkaline batteries Instrument quality certificate Electrode quality certificate Instruction manual Note Save all packing material...

Page 5: ...pH calibration at one or two points The FC10483 is a specially designed pH electrode intended for the direct determination of pH in wine production from freshly pressed fruit juice all the way to the final fermented product The FC10483 pH probe features a refillable double junction reference electrode that uses a KCl electrolyte an open junction with Clogging Prevention System CPSTM technology a b...

Page 6: ...rized buffer sets standard or NIST Selectable temperature unit C or F Electrode condition indicator mV of pH measurement for electrode check FC10483 dedicated pH with integrated temperature sensor Probe quick connect system Battery life indication and low battery detection Keystroke confirmation tone Auto off function Waterproof casing IP67 MAIN FEATURES ...

Page 7: ...intselectablebufferset Standard 3 00 7 01 10 01 or NIST 3 00 6 86 9 18 Probe included FC10483 preamplified pH and temperature probe with glass body DIN connector and 1 m 3 3 cable Battery type life 1 5V AAA 3 pcs approx 1400 hours of continuous use Auto off user selectable after 8 min 60 min or disabled Environment 0 to 50 C 32 to 122 F RH max 100 Meter dimensions 154 x 63 x 30 mm 6 1 x 2 5 x 1 2 ...

Page 8: ... 11 pH calibration buffer s used OPERATIONAL GUIDE Each meter is supplied with batteries Before using the meter for the first time open the battery compartment and insert batteries observing the polarity see Battery Replacement CONNECTING THE ELECTRODE Withthemeterturnedoff connecttheFC10483probetothe DIN socket on the bottom of the meter by aligning the pins and pushing in the plug firmly Remove ...

Page 9: ... If the probe is not connected the message NO PROBE appears alternatively on the secondary LCD with blinking on the first LCD line If the probe is not compatible WRONG PROBE message appears alternatively on the secondary LCD with blinking on the first LCD line If the readings are out of range the nearest range limits are displayed blinking e g 2 00 pH 5 0 C SELECTING MEASUREMENT RANGE While in mea...

Page 10: ...e tag on the LCD disappears The LCD displays the pH value automatically compensated for temperature on the primary LCD while the secondary LCD displays the sample temperature If measurements are taken in different samples successively rinse the probe tip thoroughly in distilled or deionized water to eliminate cross contamination Press any button to resume active measurements ENTERING CALIBRATION M...

Page 11: ... displaying pH 7 01 USE and tag blink ing or pH 6 86 USE for NIST Follow directions for single and two point calibration below Single point calibration 1 Place the probe in any buffer from the selected buffer set The meter will automatically recognize the buffer value 2 If the buffer is not recognized or the calibration offset is out of the accepted range WRONG is displayed 3 If the buffer is reco...

Page 12: ...better accuracy it is always recommended to carry out a two point calibration After the calibration procedure has been completed the tag is turned on together with the calibrated points Exiting calibration and resetting default values After entering the calibration mode and before the first point is accepted it is possible to quit the procedure and return to the last calibration data by pressing t...

Page 13: ...ere is still only 1 bar or 1 bar blinking replace the probe Sensor check Setting the meter to pH mV range the user can check the sensor status at any time The offset value is the reading in pH 7 01 buffer 25 C 77 F If this reading is outside the range 30 mV the electrode is considered very poor The slope value of the sensor is the difference between readings in pH 7 01 and in pH 3 00 buffers When ...

Page 14: ... navigate through the auto off choices 8 minutes 8 default value 60 minutes 60 or disabled Press to confirm and to enter the BEEP selection To switch the beep tone ON or OFF press the SET button press to confirm and to enter the calibration buffer selection pH 7 01 BUFF To change the buffer set the meter will show the current buffer set pH 7 01 BUFF for standard buffer set 3 00 7 01 10 01 or pH 6 ...

Page 15: ...to warn the user Battery Error Prevention System BEPS If the battery is too weak 0 the display shows bAtt DEAD for a few seconds then the meter powers off Immediately replace the batteries with new ones The batteries are accessed by opening the battery cover on the back of the instrument Remove protective boot if present Replace the three 1 5V AAA alkaline batteries located in the battery compartm...

Page 16: ...L pH 9 18 buffer solution 500 mL HI7010L pH 10 01 buffer solution 500 mL HI70300L pH electrode storage solution 500 mL HI7061L pH electrode cleaning solution 500 mL HI70635L Cleaning solution for wine deposits 500 mL HI70636L Cleaning solution for wine stains 500 mL HI7082L pH electrode fill solution 500 mL HI710029 Blue silicon rubber boot HI710142 Black carrying case for HI991XX portable instrum...

Page 17: ... the reference junctions are flowing Set aside for storage Top off the electrolyte filling solution using HI7082 solution If the bulb and or junction are dry soak the electrode in HI70300 Storage Solution for at least 30 minutes Rinse with water Calibrate before using STORAGE To ensure a quick response the glass bulb and the junction should be kept moist and not allowed to dry Code and serial numb...

Page 18: ... of purchase or go to www hannainst com CERTIFICATION ELECTRODE MAINTENANCE Replace protective cap with a few drops of HI70300 Storage Solution Follow PREPARATION above before taking measurements Replace the fill hole cover Note Never store the electrode in distilled water PERIODIC MAINTENANCE Inspect the electrode for any scratches or cracks If any present replace the electrode Rinse off any salt...

Page 19: ...ed by the warranty you will be notified of the charges incurred If the meter is to be returned to Hanna Instruments first obtain a Returned Goods Authorization number from the Technical Service department and then send it with shipping costs prepaid When shipping any meter make sure it is properly packed for complete protection Warranty Before using this product make sure it is entirely suitable f...

Page 20: ...Printed in ROMANIA MAN99111 World Headquarters Hanna Instruments Inc Highland Industrial Park 584 Park East Drive Woonsocket RI 02895 USA www hannainst com ...