53
TROUBLESHOO
TING / ERROR MESSAGES
15. TROUBLESHOOTING / ERROR MESSAGES
HI981954
displays error messages to aid in troubleshooting. Warnings are displayed for non-critical
issues. Errors are displayed for critical issues. See CALIBRATION section for messages during calibration.
“Log space full” is displayed when the meter memory is full and additional
data cannot be logged.
Delete one or more lots from the meter.
“Power fault. Check the probe cable” is displayed when powering up the
meter with a probe connected. It is triggered if the meter detects a high load
on the probe connection.
Check the probe cable. If the problem persists, contact your local Hanna
Instruments Office.
“Language data not available!” is displayed when powering up the meter if
the language file is not loaded.
Restart the meter. If the problem persists, contact your local Hanna
Instruments Office.
‘’Dead meter battery!’’ is displayed if the meter batteries are too low to
power the meter. The meter will automatically turn off.
Connect the charger if using rechargeable battery or replace the alkaline
batteries to continue.
‘’User data corrupted!’’ is displayed when powering up and the user data
stored on meter is corrupted.
Restart the meter. If the problem persists, contact your local Hanna
Instruments Office.
“Incompatible Probe” is displayed when the connected probe is not
compatible with the meter.
Replace the probe.
“Incompatible sensor detected!” is displayed when the connected sensor is
not compatible with probe or meter.
Replace the sensor before continuing.
‘’Warning x’’: Any other warning that appears at power-on is identified
using a numeric code. Restart the meter. If the problem persists, contact
your local Hanna Instruments Office. Some meter/probe features can be
accessed but with no guarantee.
‘’Errors x’’: Any critical errors that appear are identified using a numeric
code, and the meter is automatically switched off. Contact your local Hanna
Instruments Office.