7
* The 0.001 µS/cm EC range and 0.1 MΩ
•
cm Resistivity range are not available with the
1.5 m cable probe.
EC
Range
0 to 400 mS/cm
(shows values up to 1000 mS/cm)
Actual conductivity 1000 mS/cm
0.001 to 9.999 µS/cm*
10.00 to 99.99 µS/cm
100.0 to 999.9 µS/cm
1.000 to 9.999 mS/cm
10.00 to 99.99 mS/cm
100.0 to 1000.0 mS/cm (autoranging)
Resolution
0.001 µS/cm* / 0.01 µS/cm / 0.1 µS/cm
0.001 mS/cm / 0.01 mS/cm / 0.1 mS/cm
Accuracy
±1% of reading (±0.01 µS/cm or 1 digit whichever greater)
Resistivity
Range
1.0 to 99.9 Ω
•
cm
100 to 999 Ω
•
cm
1.00 to 9.99 KΩ
•
cm
10.0 to 99.9 KΩ
•
cm
100 to 999 KΩ
•
cm
1.00 to 9.99 MΩ
•
cm
10.0 to 100.0 MΩ
•
cm* (autoranging)
Resolution
0.1 Ω
•
cm / 1 Ω
•
cm / 0.01 KΩ
•
cm / 0.1 KΩ
•
cm /
1 KΩ
•
cm
0.01 MΩ
•
cm / 0.1 MΩ
•
cm*
Accuracy
±1% of reading (±10 Ω
•
cm or 1 digit whichever greater)
TDS
Range
0.00 to 99.99 ppm
100.0 to 999.9 ppm
1.000 to 9.999 g/L
10.00 to 99.99 g/L
100.0 to 400.0 g/L (autoranging)
Resolution
0.01 ppm / 0.1 ppm / 0.001 g/L / 0.01 g/L / 0.1 g/L
Accuracy
±1% of reading (±0.05 ppm or 1 digit whichever greater)
SPECIFICA
TION