
56
Data transmission from the instrument to the PC can be done with the
HI92000
Windows
®
compatible
software (optional).
HI92000
also offers graphing and on‑line help feature.
Data can be exported to the most popular spreadsheet programs for further analysis.
To connect your instrument to a PC, use an
USB
cable connector. Make sure that your instrument
is switched off and plug one connector to the instrument
USB
socket and the other to the serial or
USB port of your PC.
Note: If you are not using Hanna Instruments HI92000 software, please see the following
instructions.
SENDING COMMANDS FROM PC
It is also possible to remotely control the instrument with any terminal program. Use an
USB
cable
in accordance with the model to connect the instrument to a PC, start the terminal program and set
the communication options as follows: 8, N, 1, no flow control.
COMMAND TYPES
To send a command to the instrument follow the next scheme:
<command prefix> <command> <CR>
where:
<command prefix> is a selectable 16 ASCII character
<command> is the command code.
Note: Either lowercase or capital letters can be used.
PC INTERF
ACE
SIMPLE COMMANDS
KF1
Is equivalent to pressing
functional key 1
KF2
Is equivalent to pressing
functional key 2
KF3
Is equivalent to pressing
functional key 3
RNG
Is equivalent to pressing
RANGE
MOD
Is equivalent to pressing
MODE
CAL
Is equivalent to pressing
CAL
UPC
Is equivalent to pressing the
UP
arrow key
DWC
Is equivalent to pressing the
DOWN
arrow key
RCL
Is equivalent to pressing
RCL
SET
Is equivalent to pressing
SETUP
CLR
Is equivalent to pressing
CLR
OFF
Is equivalent to pressing
OFF