59
11 ‑ Resistivity range
12 ‑ TDS range
13 ‑ USP range
14 ‑ NaCl % range
15 ‑ Salinity, Seawater range
16 ‑ Salinity, Seawater PSU range
• Creation time (12 chars)
• Temperature source (1 char): 0 ‑ probe, 1 ‑ manual entry
• Temperature compensation: 0 ‑ NOTC, 1 ‑ linear TC, 2 ‑ non linear TC
• Beep status (1 char): 0‑ off, 1 ‑ on
• Temperature unit (1 char): 0 ‑ °C, 1 ‑ °F
• Backlight (2 chars)
• Contrast (2 chars)
• Auto light off (3 chars)
• Auto power off (3 chars)
• Baud rate (5 chars)
• Instrument ID (4 chars)
• EC alarm time out interval (2 chars)
• NaCl alarm time out interval (2 chars)
• Logging interval (5 chars)
• Calibration range check (1 char): 0 ‑ off; 1 ‑ on
• TDS factor (6 chars)
• Temperature coefficient (6 chars)
• Reference temperature (6 chars)
• Cable correction resistor (5 chars)
• Cable correction capacitance (4 chars)
• EC fixed range number (1 char)
• Resistivity fixed range number (1 char)
• TDS fixed range number (1 char)
• Language ID (3 chars)
• Cell Constant (7 chars)
PRFxx
Request profile “xx” information
The answer string contains:
10 ‑ EC range
PC INTERF