INSTRUCTION MANUAL

HI96722

Cyanuric Acid ISM

2

3-4

13•

When the correct year has been set, press 

ZERO/CFM

 or 

READ /TIMER

 to confirm. 

Now the display will show the month 
blinking.

14•

Press 

GLP/

 to edit the desired month 

(01-12). If the key is kept pressed, the 
month number is automatically increased.

15•

When the correct month has been set, press 

ZERO/CFM

 or 

READ /TIMER

 to confirm. 

Now the display will show the day blinking.

16•

Press 

GLP/

 to edit the desired day    

(01-31). If the key is kept pressed, the day 
number is automatically increased.

Note: It is possible to change the editing 

from day to year and to month by pressing  

READ

/TIMER.

17•

Press 

ZERO/CFM

 

to save the calibration 

date.

18•

The instrument displays “

Stor

” for one 

second and the calibration is saved.

19•

The instrument will return automatically to 

measurement mode by displaying dashes 
on the LCD.

LAST CALIBRATION DATE

1•

 Press 

GLP/

 to enter GLP mode. The 

calibration month and day will appear 
on the main display and the year on the 
secondary display.

2•

  If no calibration was performed, the factory 

calibration message, 

“F.CAL”

 will appear on 

the main display and the instrument returns 
to measurement mode after three seconds. 

12-13

 or

14-15

 or

16

17

18

1

2

Calibration

Note: It is possible to interrupt the calibration 

procedure at any time by pressing CAL CHECK 

or ON/OFF keys. 

1•

  Turn the meter on by pressing 

ON/OFF

.

2•

  When the beeper sounds briefly and the 

LCD displays dashes, the meter is ready.

3•

  Press and hold 

CAL CHECK 

for three 

seconds to enter calibration mode. The 
display will show 

“CAL”

 during calibration 

procedure. The blinking 

“ZERO”

 asks for 

instrument zeroing.

4•

  Place the 

CAL Check™

 Standard  

HI96722-11

 Cuvette A into the cuvette 

holder and ensure that the notch on the 
cap is positioned securely into the groove.

5•

 Press 

ZERO/CFM

 and the lamp, cuvette 

and detector icons will appear on the 
display, depending on the measurement 
phase.

6•

  After a few seconds the display will 

show     

“-0.0-”

. The meter is now zeroed 

and ready for calibration. The blinking 

“READ”

 asks for reading calibration 

standard.

7•

  Remove the cuvette.

8•

  Place the 

CAL Check™

  Standard                

HI96722-11

 Cuvette B into the holder 

and ensure that the notch on the cap is 
positioned securely into the groove.

9•

 Press 

READ /TIMER

 and the lamp, 

cuvette and detector icons will appear on 
the display, depending on the measure-
ment phase.

10•

 The instrument will show for three seconds 

the 

CAL Check™

 standard value.

Note: If the display shows “STD HIGH”, the 

standard value was too high. If the display 

shows “STD LOW”, the standard value was too 

low. Verify that both CAL Check™ Standard  

HI96720-11 Cuvettes, A and B are free 

from fingerprints or dirt and that they are 

inserted correctly.

11•

Then the date of last calibration (e.g.: 

“01.08.2009”) appears on the display, 

or “01.01.2009” if the factory calibration 

was selected before. In both cases the year 

number is blinking, ready for date input.

12•

Press 

GLP/

 to edit the desired year 

(2009-2099). If the key is kept pressed, 
the year number is automatically in-
creased. 

9

3

4

5-6

9

10-11

HI96722

HI96722

3

Validation

1•

  Turn the meter on by pressing 

ON/OFF

.

2•

  When the beeper sounds briefly and the 

LCD displays dashes, the meter is ready.

3•

  Place the 

CAL Check™

 Standard 

HI96722-11

 Cuvette  A into the holder 

and ensure that the notch on the cap is 
positioned securely into the groove.

4•

 Press 

ZERO/CFM

 and the lamp, cuvette 

and detector icons will appear on the 
display, depending on the measurement 
phase.

5•

  After a few seconds the display will show      

“-0.0-”

. The meter is now zeroed and 

ready for validation.

6•

  Remove the cuvette.

7•

  Place the 

CAL Check™

  Standard                

HI96722-11

 Cuvette B into the holder 

and ensure that the notch on the cap is 
positioned securely into the groove.

8•

 Press 

CAL CHECK

 key and the lamp, 

cuvette and detector icons together with 

“CAL CHECK”

 will appear on the display, 

depending on the measurement phase.

9•

  At the end of the measurement the display 

will show the validation standard value.  
The reading should be within specifica-
tions as  reported on the 

CAL Check™

 

Standard  Certificate. If the value is found 
out of specifications, please check that the 
cuvettes are free of fingerprints, oil or dirt 
and repeat validation. If results are still 
found out of specifications then recalibrate 
the instrument.

4-5

7

8

Warning:

 do not validate or calibrate the instrument with standard 

solutions other than the Hanna Instruments 

CAL Check™

 Standards, otherwise  

erroneous results will be obtained.
For accurate validation and calibration results, please perform tests at room 
temperature (18 to 25 °C; 64.5 to 77.0 °F).

VALIDATION AND CALIBRATION PROCEDURES

 

Use the Hanna Instruments CAL Check™ cuvettes (see 

“Accessories”) to validate or calibrate instruments.

Validation

 

Calibration

 

In GLP mode, the last calibration date can be verified and the factory calibra-
tion can be restored.

GLP

Last Calibration 
Date

 

3•

Press 

ZERO/CFM

 to restore the factory 

calibration or press 

GLP/

 again to abort 

factory calibration restore. 

4•

The instrument briefly notifies 

“donE” 

when 

restores factory calibration and returns to 
measurement mode.

Factory Calibration 
Restore

To save the battery, the instrument shuts down after 10 minutes of non-use in 
measurement mode and after 1 hour of non-use in calibration mode.

 

If a valid measurement was displayed before auto-
shut off, the value is displayed when the instrument 
is switched on. The blinking 

“ZERO” 

means that a 

new zero has to be performed.
One fresh battery lasts for around 750 measurements, depending on the 
light level.
The remaining battery capacity is evaluated at the instrument startup and 
after each measurement.
The instrument displays a battery indicator with three levels as follows:
•  3 lines for 100 % capacity
•  2 lines for 66 % capacity
•  1 line for 33 % capacity
•  Battery icon blinking if the capacity is under 10 %.
If the battery is empty and accurate measurements can’t be taken any more, 
the instrument shows 

“dEAd bAtt”

 and turns off.

To restart the instrument, the battery must be replaced with a fresh one.
To replace the instrument’s battery, follow the steps:
•  Turn the instrument off by pressing 

ON/OFF

.

•  Turn the instrument upside down and remove the battery cover by turning 

it counterclockwise.

•  Extract the battery from its location and replace it with a fresh one.
•  Insert back the battery cover and turn it clockwise to close.

BATTERY MANAGEMENT

3''

Before using these products, make sure that they are entirely suitable for 
your specific application and for the environment in which they are used.
Operation of these instruments may cause unacceptable interferences to 
other electronic equipments, this requiring the operator to take all necessary 
steps to correct interferences. 
Any variation introduced by the user to the supplied equipment may degrade 
the instrument’s EMC performance. 
To avoid damages or burns, do not put the instrument in microwave oven. 
For yours and the instrument safety do not use or store the instrument in  
hazardous environments. 

RECOMMENDATIONS FOR USERS

Factory Calibration Restore

It is possible to delete the calibration and restore 
factory calibration.

1•

Press 

GLP/

 to enter GLP mode.

 

2•

Press 

READ /TIMER

 to enter in the factory 

calibration restore screen. The instrument asks 
for confirmation of user calibration delete.  

Summary of Contents for HI96722

Page 1: ...INSTRUCTION MANUAL HI96722 Cyanuric Acid ISM ...

Page 2: ...n Manual Quality Certificate Note Save all packing material until you are sure that the instrument works correctly Any defective item must be returned in its original packing PRELIMINARY EXAMINATION For more details about spare parts and accessories see Accessories SPECIFICATIONS Range 0 to 80 mg L Resolution 1 mg L Accuracy 25 C 77 F 1 mg L 15 of reading Light source Tungsten lamp Light Detector ...

Page 3: ...own prior to measurement In GLP mode press to view the next screen 5 ON OFF key to turn the meter on and off 6 Liquid Cristal Display LCD 7 Cuvette alignment indicator 8 Cuvette holder 1 The measuring scheme lamp cuvette detector appears during different phases of zero or reading measurement 2 Error messages and warnings 3 The battery icon indicates the charge state of the battery 4 The hourglass ...

Page 4: ...25 mL mark with the sample add the content of one packet of HI93722 0 reagent and stir gently to mix 9 Fill a second cuvette with 10 mL of the reacted sample up to the mark and replace the cap 10 Place the second cuvette into the holder and ensure that the notch on the cap is positioned securely into the groove 11 Press and hold READ TIMER for three seconds The display will show the count down pri...

Page 5: ...or zeroing the meter Under range A blinking 0 0 indicates that the sample absorbs less light than the zero reference Check the procedure and make sure you use the same cuvette for reference zero and measurement Over Range A flashing value of the maximum concentration indicates an over range condition The concentration of the sample is beyond the programmed range dilute the sample and re run the te...

Page 6: ...t covered by the warranty you will be notified of the charges incurred If the instrument is to be returned to Hanna Instruments first obtain a Returned Goods Authorization Number from the Customer Service Depart ment and then send it with shipment costs prepaid When shipping any instrument make sure it is properly packaged for complete protection To validate your warranty fill out and return the e...

Page 7: ... icons together with CAL CHECK will appear on the display depending on the measurement phase 9 At the end of the measurement the display will show the validation standard value The reading should be within specifica tions as reported on the CAL Check Standard Certificate If the value is found out of specifications please check that the cuvettes are free of fingerprints oil or dirt and repeat valid...

Page 8: ...or reading calibration standard 7 Remove the cuvette 8 Place the CAL Check Standard HI96722 11 Cuvette B into the holder and ensure that the notch on the cap is positioned securely into the groove 9 Press READ TIMER and the lamp cuvette and detector icons will appear on the display depending on the measure ment phase 10 The instrument will show for three seconds the CAL Check standard value Note I...

Page 9: ...bration is saved 19 The instrument will return automatically to measurement mode by displaying dashes on the LCD LAST CALIBRATION DATE 1 Press GLP to enter GLP mode The calibration month and day will appear on the main display and the year on the secondary display 2 If no calibration was performed the factory calibration message F CAL will appear on the main display and the instrument returns to m...

Page 10: ... and accurate measurements can t be taken any more the instrument shows dEAd bAtt and turns off To restart the instrument the battery must be replaced with a fresh one To replace the instrument s battery follow the steps Turn the instrument off by pressing ON OFF Turn the instrument upside down and remove the battery cover by turning it counterclockwise Extract the battery from its location and re...

Reviews: