background image

1. INTRODUCTION

 ............................................................................................................................................................

2. UNPACKING

 ..................................................................................................................................................................

3. SAFETY MEASURES

.......................................................................................................................................................

4. INSTALLATION

..............................................................................................................................................................

4.1. AUTOSAMPLER FRONT VIEW ..........................................................................................................................................
4.2. AUTOSAMPLER REAR VIEW ............................................................................................................................................
4.3. AUTOSAMPLER SIDE VIEW ............................................................................................................................................

5. AUTOSAMPLER ASSEMBLY

 ...........................................................................................................................................

5.1. BURETTE ASSEMBLY .....................................................................................................................................................

5.1.1. DISASSEMBLING THE BURETTE .............................................................................................................................
5.1.2. ASSEMBLING THE BURETTE ..................................................................................................................................

5.2. ATTACHING THE CABLE CARRIER ....................................................................................................................................
5.3. SETTING UP DISPENSER HEAD .......................................................................................................................................
5.4. CONNECTING PUMP TUBES ...........................................................................................................................................
5.5. INSTALLING THE TRAY ...................................................................................................................................................
5.6. ELECTRICAL CONNECTIONS ............................................................................................................................................

6. OPERATION

 ..................................................................................................................................................................

6.1. STATUS LIGHTS ............................................................................................................................................................
6.2. CONTROL PANEL ...........................................................................................................................................................

7. MAINTENANCE

 .............................................................................................................................................................

7.1. REPLACING PERISTALTIC PUMP TUBING ..........................................................................................................................
7.2. REMOVING A PUMP ......................................................................................................................................................
7.3. ADDING A PUMP ..........................................................................................................................................................
7.4 UPGRADING FIRMWARE .................................................................................................................................................

8

TECHNICAL

 

SPECIFICATION

 .........................................................................................................................................

9

AUTOSAMPLER

 

COMPONENTS

 .....................................................................................................................................

5

6

14

7

8

8

14

9

15

10

16

12

18

5

7

14

7

8

8

15

9

16

11

17

13

19

TABLE OF CONTENTS

Summary of Contents for HI922

Page 1: ...INSTRUCTION MANUAL AUTOSAMPLER HI922 ...

Page 2: ......

Page 3: ... the autosampler s functionality see HI932 Automatic Potentiometric Titrator Manual Chapter 12 If you need additional technical information do not hesitate to e mail us at tech hannainst com or view our worldwide contact list at www hannainst com for the nearest Hanna Instruments representative Dear Customer 2018 Hanna Instruments All rights are reserved Reproduction in whole or in part is prohibi...

Page 4: ...URETTE 5 2 ATTACHING THE CABLE CARRIER 5 3 SETTING UP DISPENSER HEAD 5 4 CONNECTING PUMP TUBES 5 5 INSTALLING THE TRAY 5 6 ELECTRICAL CONNECTIONS 6 OPERATION 6 1 STATUS LIGHTS 6 2 CONTROL PANEL 7 MAINTENANCE 7 1 REPLACING PERISTALTIC PUMP TUBING 7 2 REMOVING A PUMP 7 3 ADDING A PUMP 7 4 UPGRADING FIRMWARE 8 TECHNICAL SPECIFICATION 9 AUTOSAMPLER COMPONENTS 5 6 14 7 8 8 14 9 15 10 16 12 18 5 7 14 7 ...

Page 5: ... single box containing ITEM QUANTITY Autosampler 1 pc Control Panel 1 pc Tray Locking Screw 1 pc Titrant Dispensing Tube 1 pc Electrode Holder 1 pc Communication Cable 1 pc BNC Extension Cable 1 pc Temperature Sensor 1 pc Beaker Tray Assembly 1 pc Plastic Beaker 20 pcs Stir Bars 25 mm x 7 mm 10 pcs USB Flash Drive 1 pc Instruction Manual 1 pc Quality Certificate 1 pc For a complete list of Autosam...

Page 6: ...ns Severe vibrations Direct sunlight Atmospheric relative humidity above 95 non condensing Environment temperatures below 10 C and above 40 C Explosion hazards 6 Have the autosampler serviced by qualified service personnel only 7 Keep hands and objects clear of moving parts during operation In an emergency situation the autosampler can be stopped immediately by pressing either of the illuminated e...

Page 7: ...7 4 INSTALLATION 4 1 AUTOSAMPLER FRONT VIEW 4 2 AUTOSAMPLER REAR VIEW INSTALLATION ...

Page 8: ...ted on the left side and the dispensing tube is mounted on the right side of the burette To remove the dispensing tube follow these steps Remove the blue tube protector A by sliding it off the clear titrant tubing Remove the tube lock B from the burette holder Turn the fitting C counterclockwise to remove it from the burette holder Slide the clear titrant tubing through the fitting INSTALLATION AS...

Page 9: ...Straighten out tube up into the vertical position to enter the highest cut of the fitting C Attach the tube lock fitting D Attach the blue tube protector E by sliding it over the clear titrant tubing the protector will be positioned in the tube lock fitting 5 2 ATTACHING THE CABLE CARRIER To attach the cable carrier remove the three 3 screws from the top rear panel A Align the screw holes on the c...

Page 10: ...flat head screwdriver Insert electrode temperature sensor stirrer and aspiration tube optional into the dedicated holes in the electrode holder Push them down until they are in a stable position Insert burette and auxiliary tubing into the appropriate holders Close clips on cable carrier once all cables and tubing are positioned ASSEMBLY ...

Page 11: ...eft side of the peristaltic pump Connect the reagent container tubing to the right side of the peristaltic pump For Waste Aspiration Operation Connect the waste container tubing to the left side of the peristaltic pump Connect the aspiration tubing to the right side of the peristaltic pump ASSEMBLY ...

Page 12: ...RAY Place the tray on the turntable with Beaker 1 under the dispenser Make sure the turntable mounting pins are aligned with the tray Lock the tray in place with the locking screw Do not overtighten the locking screw ASSEMBLY ...

Page 13: ...CTIONS Connect the titrator to Titrator connector Connect the control panel to Control Panel connector Connect the overhead stirrer optional to Stirrer connector Connect a USB barcode reader optional to the USB port ASSEMBLY ...

Page 14: ...or user action Red flashing Error Red steady Firmware corrupted Autosampler initializing first 3 seconds after power on 6 2 CONTROL PANEL The control panel displays the current status of the autosampler and can be used to manually control the autosampler while it is idle green steady status lights Move the dispenser head up down Rotate the turnable clockwise counterclockwise Power the auxiliary pu...

Page 15: ...ic cover off of the pump 1 Remove the plastic rotor and tubing 2 Thoroughly grease the outside of the replacement tube and the rotor s rollers Wrap the replacement tubing around the rotor and press into the pump housing 3 The tubing in contact with the rollers will need to be compressed with a finger in order to fit in the housing Replace the clear plastic cover MAINTENANCE ...

Page 16: ... the pump out of the autosampler tower If the pump s bay is no longer being used replace the side and back covers and fasten with the 4 screws 7 3 ADDING A PUMP Peristaltic pumps can only be connected to bay 1 2 and 3 Membrane pump can only be connected to bay 4 Remove the 4 screws holding side and rear covers of the pump bay Insert the pump into the rear of the autosampler tower Once inside the t...

Page 17: ...tch but leave the cable connected to the HI932 Titrator 3 Insert the USB flash drive with upgrade file in the HI922 USB port 4 While holding down both emergency stop buttons turn on the HI922 Autosampler Keep the buttons pressed until the status lights turn yellow The control panel should display the message Upgrading FW please wait After this message appears release the buttons and wait for the u...

Page 18: ...s x 100 mL with Built in RFID Beakers ASTM short form glass beakers 100 150 mL HI920 060 150 mL Plastic beakers HI920 053 100 mL Plastic beakers Control Panel Buttons for manual operation of tray Manual operation of peristaltic or diaphragm pumps 2 line backlight display with status information Enclosure Material ABS plastic and steel Electrode Holder Material ABS plastic Tray Material ABS plastic...

Page 19: ...0 11660 20 pcs HI920 060 Plastic Beaker for HI920 11853 20 pcs HI920 053 Electrode Holder HI920 310 Temperature Sensor HI7662 AW USB Flash Drive HI920 901 Titrant Dispensing Tube 1 5 m HI920 281 Tray Locking Screw HI920 960 Reference Extension Cable 1 m HI920 932 Autosampler HI922 XYZ Control Panel HI920 922 Communication Cable HI920 933 HI932 to HI921 HI922 BNC Extension Cable 1 m HI920 931 ...

Page 20: ...e for Autosampler peristaltic pump 3 pcs HI920 204 Replacement Propellers 3 pcs HI930302 High Chemical Resistance Propellers 3 pcs HI930303 Cable Chain HI920 320 25 mm x 7 mm Stir Bars 10 pcs HI731319 Peristaltic pump with aspiration tubing HI920 104 Membrane Pump with tubing HI920 113 TYGON Tube 5 m HI920 290 Overhead Stirrer 3 propellers HI930301 Peristaltic Pump with dispensing tubing HI920 103...

Page 21: ...te collection point for the recycling of electrical and electronic equipment which will conserve natural resources Ensuring proper product and battery disposal prevents potential negative consequences for the environment and human health For more information contact your city your local household waste disposal service the place of purchase or go to www hannainst com Certification MAN922 10 18 ...

Page 22: ...e problem If the repair is not covered by the warranty you will be notified of the charges incurred If the instrument is to be returned to Hanna Instruments first obtain a Returned Goods Authorization RGA numberfromtheTechnical Service department and then send it with shipping costs prepaid When shipping any instrument make sure it is properly packed for complete protection Warranty Before using t...

Page 23: ......

Page 24: ...and Industrial Park 584 Park East Drive Woonsocket RI 02895 USA www hannainst com Local Office Hanna Instruments Inc Highland Industrial Park 584 Park East Drive Woonsocket RI 02895 USA Phone 800 426 6287 Fax 401 765 7575 e mail tech hannainst com ...

Reviews: