WARNING!
Turn off the autosampler and disconnect from the titrator before adding or removing pumps! Failure to do so could cause
damage to the pump and/or autosampler.
7.2. REMOVING A PUMP
• Remove the 4 screws holding the pump in the autosampler tower (1).
• Pull the pump out (2) of the tower slightly so that the pump’s side cover can be removed (3).
• Disconnect the pump's electrical connection by holding down the connector release tab (4) while pulling the connector away
from the circuit board (5).
• Pull the pump out of the autosampler tower.
• If the pump's bay is no longer being used, replace the side and back covers and fasten with the 4 screws.
7.3. ADDING A PUMP
Peristaltic pumps can only be connected to bay #1, #2, and #3. Membrane pump can only be connected to bay #4.
• Remove the 4 screws holding side and rear covers of the pump bay.
• Insert the pump into the rear of the autosampler tower.
• Once inside the tower, pull the pump's electrical connector out of the side of the tower. Push the connector into the circuit board
with the push-tab facing the outside of the tower.
• Replace the side cover. You will need to pull the pump out of the tower slightly.
• Fasten the pump in place with 4 screws.