35
PU
m
P C
ALIB
rA
TION
Pr
OCEDU
rE
Notes:
If temperature probe is not connected, Manual Temperature Compensation is used and MTC
appears on the right side of the screen. If Automatic Temperature Compensation is in use, the
ATC appears on the right side of the screen.
If an erroneous situation is encountered during the calibration, an error message is displayed and
the calibration can be restarted by pressing Restart. Prepare a new standard, rinse the electrode,
temperature probe and dosing tip and try again.
If the calibration doesn’t complete and the max titrant volume of titrant is reached, an error
message will be displayed. The calibration can be restarted by pressing Restart. Prepare a new
standard, rinse electrode, temperature probe and dosing tip and try again.
This error message appears when the pH reading exceeds the acceptable input limits (-2.00 <
pH > 16.00).
This screen appears when the stirrer is not working properly. Check the stir bar and beaker
content. Press Restart to try again.
This error message appears when the pump is not working properly. Check the tubing, valve and
syringe. Press Restart to try again.