33
PU
m
P C
ALIB
rA
TION
Pr
OCEDU
rE
14. PUMP CALIBRATION PROCEDURE
The calibration of the pump must be performed every time the syringe, pump tubing, the titrant bottle
or the pH electrode is changed. A pump calibration is recommended before each set of titrations, after
the titrator is left idle for several hours, or once daily.
Note:
A pump calibration needs to be done in the desired range and sample size. Pump calibration
is independent of the sample type.
• Press
MENU
, select
Setup
and select the corresponding range according to the table below:
Unit Low Range 20 or 50 High Range 20
%l.a.
0.01 to 0.20
0.1 to 2.0
°SH
0.4 to 8.9
4.4 to 88.9
°D
1.0 to 20.0
10 to 200
°TH
1.1 to 22.2
11.1 to 222.2
• Verify the electrode has been calibrated in 8.30 pH buffer.
• Ensure the pump is primed with the correct titrant for the selected range (
HI84529-50
Low Range
20 Titrant or
HI84529-52
Low Range 50 Titrant,
HI84529-51
High Range 20 Titrant).
Sample preparation:
Use a clean pipette, add a precisely measured amount of
HI84529-55
Calibration Standard to a clean beaker as indicated below:
Low Range 20 – 1 mL
Low Range 50 – 2 mL
High Range 20 – 10 mL
Note:
Failure to use a clean pipette will result in erroneous readings.
• Fill the beaker up to the 50 mL mark with the distilled or deionized water.
• Press
CAL
key. The instrument displays the date and time of the last electrode calibration, and the
last pump calibration for the selected range.
• Press
Pump
key.
Note:
Do not place the tip into the calibration beaker, place the tip over a waste beaker. A small
amount of titrant is dispensed when the pump resets.
50 mL
25 mL
15 mL