Periodic Maintenance
Inspect the electrode and the cable. The cable used for connection to the instrument must be intact
and there must be no points of broken insulation on the cable or cracks on the electrode stem or
bulb. Connectors must be perfectly clean and dry. If any scratches or cracks are present, replace the
electrode. Rinse off any salt deposits with water.
Probe Maintenance
Refill the reference chamber with fresh electrolyte (
HI7082
). Allow the electrode to stand upright for
1 hour. Follow the Storage Procedure above.
pH Cleaning Procedure
• Wine deposits: Soak in Hanna Instruments
HI70635
cleaning solution for 15 minutes
• Wine stains: Soak in Hanna Instruments
HI70636
cleaning solution for 15 minutes
Important:
After performing any of the cleaning procedures, rinse the electrode thoroughly with
distilled water, refill the reference chamber with fresh electrolyte and soak the electrode in
HI70300 Storage solution for at least 1 hour before taking measurements.