7. SETUP MENU
The titrator’s setup menu may be accessed from the main screen (meter or titrator) by pressing the
MENU
key, then
Setup
.
A list of setup parameters will be displayed with currently configured setting.
While in the setup menu it is possible to modify the instrument’s operation parameters. The
ARROW
keys permit the user to scroll the setup parameters.
Press
HELP
to view the contextual help.
Press
ESC
to return to the main screen.
Range Setup
Use
Low
measurement range for 0.1 - 5.0 g/L.
Use
High
measurement range for 4.0 - 25.0 g/L.
To ensure a high accuracy, it is recommended to recalibrate the pump after the valve, titrant or
electrode has been changed.
Calibration Timeout
Disable or 1 to 7 days
This option is used to set the number of days before the pH calibration expired warning message is
displayed.
Press
Modify
to access the calibration timeout screen.
Use the
ARROW
keys to select the value.
Press
Select
to confirm or
ESC
to return to the setup menu without saving the changes.