• If the computed slope isn’t within the preset limits, the meter will display the message “Wrong
Slope”. If the slope is too high the symbol will be displayed. If the slope is too low the symbol
will be displayed.
• If the “Wrong Old Slope” error message is displayed, an inconsistency exists between the current
and the previous (old) calibration. Clear the previous calibration by pressing Clear and proceed with
calibration from the current calibration point. The instrument will keep all the confirmed values
during the current calibration.
• If the temperature reading is out of the defined temperature range of the buffer (0 to 45 °C), the
“Wrong Buffer Temperature” error message will be displayed, and the temperature symbol will
blink on the display. Calibration cannot be confirmed in this situation.
Notes:
To clear a previous calibration and to return to the default value, press Clear at any time
after entering calibration mode. If Clear is invoked during the first calibration point the instrument
returns to the measurement mode.
The Clear key is displayed only if a previous calibration exists.