Hanna Instruments HI8424NEW Instruction Manual Download Page 8

8

Notes

:

• To perform correct ORP measurements, the surface of the ORP electrode must

be clean and smooth.

• When not in use, the tip of the electrode should be kept moist (use 

HI 70300

storage solution) and safe from any mechanical stress which might cause
damage to the glass/platinum junction.

TEMPERATURE MEASUREMENTS

• Turn the instrument ON and press the RANGE key to select the temperature

mode.

• Make sure the temperature probe is connected to the meter.
• Dip the temperature probe into the sample, allow the reading to stabilize

(hourglass symbol turns off) and read the temperature value.

• Temperature measurements can be displayed in °C or °F units (see "Menu

selection" for details).

Notes

:

• A blinking full scale value means that the reading is out of range.
• To freeze a reading on display while in mea-

surement mode, press the HOLD key. The "HOLD"

tag will blink. The pH, mV and temperature

values are held, and the RANGE key can be

used to view the values. Press HOLD again to

return to normal mode.

• If enabled, keypresses are followed with an acoustic signal. A lower note

indicates that the key is not currently active.

• To save battery life, the meter is provided with an 

auto-off feature

, which

turns the instrument off after 20 minutes of non-use. This feature can be

disabled by the user (see "Menu selection" for details).

21

120 mm

4.7"

5mm

0.2"

5mm DIA

0.2"

"S" VERSION

HI 1330

210 mm

8.25"

8 mm

0.3"

7.5mm DIA

0.29"

"S" VERSION

HI 1331

120 mm

4.7"

12 mm

0.5"

"S" VERSION

HI 1230

HI 1083B

Glass body, micro, Viscolene, non-refillable, combination 

pH

 electrode.

Use: biotechnology, micro titration.

HI 1131B / HI 1111S

Glass body, single junction, refillable, combination 

pH

 electrode.

Use: general purpose.

HI 1230B / HI 1210S

Plastic body, double junction, gel-filled, combination 

pH 

electrode.

Use: general purpose.

HI 1330B / HI 1310S

Glass body, semimicro, single junction, refillable, combination 

pH 

electrode.

Use: laboratory.

HI 1331B / HI 1311S

Glass body, semimicro, single junction, refillable, combination 

pH 

electrode.

Use: flasks.

120 mm

4.7"

12 mm

0.5"

5 mm

0.2"

3 mm

0.12"

3.0 mm DIA

0.12"

HI 1083

120 mm

4.7"

12 mm

0.5"

9.5mm DIA

0.37"

"S" VERSION

HI 1131

[email protected]

www.ponpe.com

Summary of Contents for HI8424NEW

Page 1: ...ax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 SouthAfrica Tel 011 615 6076 Fax 011 615 8582 Taiwan Tel 886 2 2739 3014 Fax 886 2 2739 2983 Thailand...

Page 2: ...f under warranty report the model number date of purchase serial number and the nature of the problem If the repair is not covered by the warranty you will be notified of the charges incurred If the i...

Page 3: ...mV to 1 mV when readings exceed 699 9 mV The user can set the following parameters by entering a setup mode through the keyboard enabling or disabling the auto off feature enabling or disabling the a...

Page 4: ...200B FC 200S 6 Dairyproducts FC 911B FC 100B 7 Dirtywater HI 1230B HI 1210S 8 Emulsions HI 1053B HI1050S 9 Environment HI 1230B HI1210S 10 Flasks HI 1331B HI1310S 11 Foodindustrygeneraluse FC 911B FC...

Page 5: ...ze thermometer 50 0 to 150 0 C HI 98502 ChecktempF pocket size thermometer 58 0 to 302 0 F HI 710015 Shockproof rubber boot blue HI 710016 Shockproof rubber boot orange HI 710022 Spare protective case...

Page 6: ...ow the pH value automatically compensated for temperature Notes In order to take accurate pH measurements make sure that the instrument has been calibrated before use see page 9 If measurements are ta...

Page 7: ...P MEASUREMENTS Oxidation Reduction Potential ORP measurements provide a quantification of the oxidizing or reducing power of the sample tested Connect the ORP electrode optional to the BNC connector T...

Page 8: ...eturn to normal mode If enabled keypresses are followed with an acoustic signal A lower note indicates that the key is not currently active To save battery life the meter is provided with an auto off...

Page 9: ...ximately 4 cm 1 into the solution The temperature probe should be located close to the pH electrode Press RANGE to display pH measurement Press CAL to enter the calibration mode The buffer value 25 C...

Page 10: ...TION SOLUTIONS HI 70004P pH 4 01 buffer solution 20 mL sachet 25 pcs HI 7004M pH 4 01 buffer solution 230 mL bottle HI 7004L pH 4 01 buffer solution 500 mL bottle HI 8004L pH 4 01 buffer solution 500...

Page 11: ...cm 1 in the sample and stir gently for a few seconds For a faster response and to avoid cross contamination of the samples rinse the electrode tip with the solution to be tested before taking any mea...

Page 12: ...ext step At this point it is possible to set the temperature unit by selecting Unit C or Unit F Press CAL SELECT to toggle the selection and CFM SETUP to exit the menu selection mode and return to nor...

Page 13: ...ak the electrode tip excessive drift in HI 7061 solution for 30 minutes Reading fluctuates Clogged dirty junction Soak the electrode tip up and down noise or low electrolyte level in warm HI 7082 refi...

Page 14: ...eplacement should take place in a non hazardous area using a 9V Ni MH rechargeable battery Note The meter can also be powered with a normal 9V alkaline battery NEVER USE THE HI 710040 CHARGER WITH NON...

Reviews: