Hanna Instruments HI8424NEW Instruction Manual Download Page 5

5

SPECIFICATIONS

SPECIFICATIONS

SPECIFICATIONS

SPECIFICATIONS

SPECIFICATIONS

Range

-2.00 to 16.00 pH

±699.9 mV / ±1999 mV

-20.0 to 120.0°C / -4.0 to 248.0°F

Resolution

0.01 pH / 0.1 mV / 1 mV / 0.1°C / 0.1°F

Accuracy (@20°C/68°F)

±0.01 pH / ±0.2 mV / ±1 mV / ±0.4°C / ±0.8°F

Typical EMC Deviation

±0.02 pH / ±0.2 mV / ±1 mV / ±0.4°C / ±0.8°F

pH Calibration

Automatic, 1 or 2 point,

 with 3 memorized buffer values (pH 4.01, 7.01, 10.01)

Offset: ±1 pH; Slope: from 75 to 110%

Temperature Compensation

Automatic, -20 to 120°C (-4 to 248°F)

or manual without temperature probe

Probes (included)

HI 1230B

 double junction, gel-filled pH electrode

HI 7662 

temperature probe

Battery Type

1 x 9V, Ni-MH rechargeable

Battery Charge Life

Approx. 150 hours of continuous use

Battery Life

Approx. 5 years

Auto-off

After 20 minutes of nin-use or disabled (user-selectable)

Environment

0 to 50°C (32 to 122°F); RH max 100%

Dimensions

164 x 76 x 45 mm (6.5 x 3.0 x 1.8'')

Weight

180 g (6.3 oz.)

24

OTHER ACCESSORIES

HI 98501

ChecktempC

 pocket-size thermometer (–50.0 to 150.0°C)

HI 98502

ChecktempF

 pocket-size thermometer (–58.0 to 302.0°F)

HI 710015

Shockproof rubber boot, blue

HI 710016

Shockproof rubber boot, orange

HI 710022

Spare protective case

HI 710040

No-contact, inductive battery charger

HI 76405

Electrode holder

HI 7662

Temperature probe with 1 m (3.3') screened cable

HI 8427

pH/ORP electrode simulator with 1 m (3.3') coaxial cable and
BNC connector

HI 931001

pH/ORP electrode simulator with LCD, 1 m (3.3') coaxial cable
and BNC connector

120 mm

4.7"

12 mm

0.5"

HI 4430

"S" VERSION

HI 4430B / HI 4410S

Plastic body, gel-filled, combination gold 

ORP

 electrode.

Use: general purpose.

Consult the HANNA General Catalog for a complete and wide selection of

electrodes.

EXTENSION CABLES FOR SCREW-TYPE ELECTRODES (SCREW
TO BNC CONNECTOR)

HI 7855/1

Extension cable 1 m (3.3') long

HI 7855/3

Extension cable 3 m (9.9') long

HI 7855/5

Extension cable 5 m (16.5') long

HI 7855/10

Extension cable 10 m (33') long

HI 7855/15

Extension cable 15 m (49.5') long

CONNECT TO
SCREW TYPE

ELECTRODES

CONNECT TO THE

BNC SOCKET

OF THE METER

HI 7855 SERIES CABLE CONNECTORS

CONNECTOR AND 3.0 mm (0.12") CABLE WITH BNC

[email protected]

www.ponpe.com

Summary of Contents for HI8424NEW

Page 1: ...ax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 SouthAfrica Tel 011 615 6076 Fax 011 615 8582 Taiwan Tel 886 2 2739 3014 Fax 886 2 2739 2983 Thailand...

Page 2: ...f under warranty report the model number date of purchase serial number and the nature of the problem If the repair is not covered by the warranty you will be notified of the charges incurred If the i...

Page 3: ...mV to 1 mV when readings exceed 699 9 mV The user can set the following parameters by entering a setup mode through the keyboard enabling or disabling the auto off feature enabling or disabling the a...

Page 4: ...200B FC 200S 6 Dairyproducts FC 911B FC 100B 7 Dirtywater HI 1230B HI 1210S 8 Emulsions HI 1053B HI1050S 9 Environment HI 1230B HI1210S 10 Flasks HI 1331B HI1310S 11 Foodindustrygeneraluse FC 911B FC...

Page 5: ...ze thermometer 50 0 to 150 0 C HI 98502 ChecktempF pocket size thermometer 58 0 to 302 0 F HI 710015 Shockproof rubber boot blue HI 710016 Shockproof rubber boot orange HI 710022 Spare protective case...

Page 6: ...ow the pH value automatically compensated for temperature Notes In order to take accurate pH measurements make sure that the instrument has been calibrated before use see page 9 If measurements are ta...

Page 7: ...P MEASUREMENTS Oxidation Reduction Potential ORP measurements provide a quantification of the oxidizing or reducing power of the sample tested Connect the ORP electrode optional to the BNC connector T...

Page 8: ...eturn to normal mode If enabled keypresses are followed with an acoustic signal A lower note indicates that the key is not currently active To save battery life the meter is provided with an auto off...

Page 9: ...ximately 4 cm 1 into the solution The temperature probe should be located close to the pH electrode Press RANGE to display pH measurement Press CAL to enter the calibration mode The buffer value 25 C...

Page 10: ...TION SOLUTIONS HI 70004P pH 4 01 buffer solution 20 mL sachet 25 pcs HI 7004M pH 4 01 buffer solution 230 mL bottle HI 7004L pH 4 01 buffer solution 500 mL bottle HI 8004L pH 4 01 buffer solution 500...

Page 11: ...cm 1 in the sample and stir gently for a few seconds For a faster response and to avoid cross contamination of the samples rinse the electrode tip with the solution to be tested before taking any mea...

Page 12: ...ext step At this point it is possible to set the temperature unit by selecting Unit C or Unit F Press CAL SELECT to toggle the selection and CFM SETUP to exit the menu selection mode and return to nor...

Page 13: ...ak the electrode tip excessive drift in HI 7061 solution for 30 minutes Reading fluctuates Clogged dirty junction Soak the electrode tip up and down noise or low electrolyte level in warm HI 7082 refi...

Page 14: ...eplacement should take place in a non hazardous area using a 9V Ni MH rechargeable battery Note The meter can also be powered with a normal 9V alkaline battery NEVER USE THE HI 710040 CHARGER WITH NON...

Reviews: