Hanna Instruments HI8424NEW Instruction Manual Download Page 13

13

HI 8424NEW 

has been accurately precalibrated for temperature at the factory.

For optimum accuracy, it is recommended to recalibrate the meter for tempera-

ture at least once a year.
Contact your dealer or the nearest HANNA instruments

®

 Customer Service

Center for more information.

HI 8424NEW 

has been accurately precalibrated for mV range at the factory.

For optimum accuracy, it is recommended to recalibrate the meter for mV

readings at least once a year.
Contact your dealer or the nearest HANNA instruments

®

 Customer Service

Center for more information.

m

m

m

m

mV CALIBRATION

V CALIBRATION

V CALIBRATION

V CALIBRATION

V CALIBRATION

TEMPERATURE CALIBRATION

TEMPERATURE CALIBRATION

TEMPERATURE CALIBRATION

TEMPERATURE CALIBRATION

TEMPERATURE CALIBRATION

16

TROUBLESHOOTING

Symptom

Problem

Solution

Slow reponse or

Dirty pH electrode

Soak the electrode tip

excessive drift

in 

HI 7061

 solution for

30 minutes

Reading fluctuates

Clogged/dirty junction

Soak the electrode tip

up and down (noise)

or low electrolyte level

in warm 

HI 7082

(refillable electrodes)

solution for one hour,
then rinse it with
distilled water
(refill with fresh
electrolyte if necessary)

Blinking full scale

Reading is out of range

value
Blinking "°C" (or "°F")

Temperature probe is not connected or broken

"WRONG" &

Calibration error

Check buffer solution or

blinking dashes

replace pH electrode

Blinking battery

Low battery level

Recharge or

symbol

replace battery

Meter shuts off

Auto-off enabled or
dead battery

Replace battery

"Clr" message

Loaded default

Perform pH calibration

pH calibration values

"Er1" and Er2"

EPROM error

Contact your dealer or

messages

HANNA Service Center

Note:

For field applications, it is always recommended to keep a conditioned spare elec-
trode handy. When anomalies cannot be resolved with simple maintenance, change
the electrode and recalibrate the meter.

[email protected]

www.ponpe.com

Summary of Contents for HI8424NEW

Page 1: ...ax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 SouthAfrica Tel 011 615 6076 Fax 011 615 8582 Taiwan Tel 886 2 2739 3014 Fax 886 2 2739 2983 Thailand...

Page 2: ...f under warranty report the model number date of purchase serial number and the nature of the problem If the repair is not covered by the warranty you will be notified of the charges incurred If the i...

Page 3: ...mV to 1 mV when readings exceed 699 9 mV The user can set the following parameters by entering a setup mode through the keyboard enabling or disabling the auto off feature enabling or disabling the a...

Page 4: ...200B FC 200S 6 Dairyproducts FC 911B FC 100B 7 Dirtywater HI 1230B HI 1210S 8 Emulsions HI 1053B HI1050S 9 Environment HI 1230B HI1210S 10 Flasks HI 1331B HI1310S 11 Foodindustrygeneraluse FC 911B FC...

Page 5: ...ze thermometer 50 0 to 150 0 C HI 98502 ChecktempF pocket size thermometer 58 0 to 302 0 F HI 710015 Shockproof rubber boot blue HI 710016 Shockproof rubber boot orange HI 710022 Spare protective case...

Page 6: ...ow the pH value automatically compensated for temperature Notes In order to take accurate pH measurements make sure that the instrument has been calibrated before use see page 9 If measurements are ta...

Page 7: ...P MEASUREMENTS Oxidation Reduction Potential ORP measurements provide a quantification of the oxidizing or reducing power of the sample tested Connect the ORP electrode optional to the BNC connector T...

Page 8: ...eturn to normal mode If enabled keypresses are followed with an acoustic signal A lower note indicates that the key is not currently active To save battery life the meter is provided with an auto off...

Page 9: ...ximately 4 cm 1 into the solution The temperature probe should be located close to the pH electrode Press RANGE to display pH measurement Press CAL to enter the calibration mode The buffer value 25 C...

Page 10: ...TION SOLUTIONS HI 70004P pH 4 01 buffer solution 20 mL sachet 25 pcs HI 7004M pH 4 01 buffer solution 230 mL bottle HI 7004L pH 4 01 buffer solution 500 mL bottle HI 8004L pH 4 01 buffer solution 500...

Page 11: ...cm 1 in the sample and stir gently for a few seconds For a faster response and to avoid cross contamination of the samples rinse the electrode tip with the solution to be tested before taking any mea...

Page 12: ...ext step At this point it is possible to set the temperature unit by selecting Unit C or Unit F Press CAL SELECT to toggle the selection and CFM SETUP to exit the menu selection mode and return to nor...

Page 13: ...ak the electrode tip excessive drift in HI 7061 solution for 30 minutes Reading fluctuates Clogged dirty junction Soak the electrode tip up and down noise or low electrolyte level in warm HI 7082 refi...

Page 14: ...eplacement should take place in a non hazardous area using a 9V Ni MH rechargeable battery Note The meter can also be powered with a normal 9V alkaline battery NEVER USE THE HI 710040 CHARGER WITH NON...

Reviews: