42
• Use
or
to select the
Calibration
option.
• Press
and use
or
to highlight
the
Edit Standard Group
option.
• Press
and use
and
to choose
the ISE standard to be included in the standard group.
• Press
/
to add/remove the selected ISE
standard to/from the standard group.
• Press
to return to
Calibration
options and to
save the changes.
Calibration Reminder
- See
Calibration
option from
pH
Setup
section.
Set Reminder Period
- See
Calibration
option from
pH
Setup
section.
Clear Calibration
- See
Calibration
option from
pH Setup
section.
Electrode Type
This option allows the user to select the desired Ion Selective Electrode used for measurements from
a list: Ammonia, Bromide, Cadmium, Calcium, Carbon Dioxide, Chloride, Cupric, Cyanide, Fluoride,
Iodide, Lead, Nitrate, Potassium, Silver, Sodium, Sulfate, Sulfide and five custom ISE. For the standard
ISE it is possible to view the Ion constants (Name, Molar Weight and Electric Charge/Slope), while for
the custom ISE all these constants can be manually set.
To set the
Electrode Type
:
• Press
while in
ISE Measure
mode.
• Press
.
• Use
or
to select the
Electrode Type
option.
• Press
and use
or
to select
the desired standard ISE or a custom one from the list.
For standard ISE:
• Press
to visualize the Ion Constants and then
press
at any time to exit Ion Constants view
mode.
ISE SETUP (HI5222 only)