12
GLP
Detailed
GLP data will be displayed on the LCD for pH Measure and ISE Measure modes only
when this option is selected: Last Calibration date and time, Offset and Slope values, Calibration
Buffers/Standards and general information regarding the buffers/standards: the calibration
temperature, temperature compensation mode, date and time. For
pH Measure
, the Electrode
Condition is also displayed on the LCD in percent.
Note: If a one point pH calibration is performed or the current calibration does not include at
least two consecutive standard buffers of pH 4.01, 7.01 (6.86) and 10.01 (9.18) buffers, the
Electrode Condition will be unknown. Electrode condition remains active for 24 hours after a
calibration.
To access the
GLP
display option:
• Press
while in
Measure
mode. The “Choose Display Configuration” message will be
displayed in the Reminder messages area.
• Press
. The instrument will display the detailed GLP data.
DISPLA
YING MODES