background image

 

3.

 

When probe is not in use, place the hard protective storage cap over glass bulb and replace fill hole 
cap. 

 

MEASUREMENT GUIDES: 

1.

 

Always clamp or support probe during use. 

2.

 

Frequent calibrations are suggested for critical pH applications.  

3.

 

To eliminate thermal or sample gradients stir buffers and samples using a magnetic stirrer. 

4.

 

The reference junction may be cleaned by lifting the PTFE sleeve. Refill sensor with fresh HI7082 
after replacing sleeve. 

 

CALIBRATION: 

1.

 

Calibration can be performed using up to five calibration buffers. For accurate measurements a 
minimum of a three - point calibration is recommended. Please note the HI10482 probe will use pH 
3.00 buffer instead of pH 4.01. 

2.

 

Enter calibration screen (using Hanna Lab App or edge®blue). 

3.

 

Place the probe in the first buffer. 

4.

 

Confirm the calibration point when the reading is stable. 

5.

 

Repeat the procedure for up to five calibration points. For details, please see edge®blu manual or 
Hanna Lab App HELP. 

 

STORAGE: 

1.

 

Rinse off the HI10482/HALO™ probe with deionized water to remove sample. 

2.

 

Blot off using a soft tissue. 

3.

 

Pour 1 cm of HI 70300 storage solution into plastic storage cap. Place probe bulb into cap and tighten 
the threaded collar. 