34
The instrument will display a lot of data including calibration standards,
offset, time and date, etc.
Use the
Arrow
keys to select the offset or calibration standards, in order to
view new information.
To see more information press
More
.
• More information regarding offset.
• More information regarding standards.
Notes:
• Standards displayed in video inverse mode are from previous
calibrations.
•
“No user calibration”
message is displayed if all calibrations
are cleared or the instrument was not calibrated in the EC range.
LAST NaCl % CALIBRATION DATA
Last NaCl calibration data is stored automatically after a successful calibration.
To view the NaCl calibration data, press
GLP
key while in NaCl measurement
mode.
The instrument will display the NaCl information: calibration date, time and
offset.