23
Wait until the reading is stable (about 5 minutes).
Note:
If the input record has an instability higher than 1
m
S, stability period
will be reset. The completion time bar will remain empty.
The “
USP Met
” message will be displayed if the USP stage 2 criteria is reached.
Press
Report
to view the USP report.
Press key to scroll the report screens.
Press
Log
to memorise report.
Notes:
• The record number will be the same as at the Stage 1 report
(informations of the same analyse).
• If the log space is full enter view logged data mode by pressing
RCL
key and free log space by deleting previously memorized records.
The “
USP Not Met
” message is displayed if the USP stage 2 criteria is not met.
Press
Report
to view report.
Press key to scroll the report screens. Press
Log
to memorise report.
Press
Stage 3
to enter in stage 3 water analyze.
The
USP Stage 3
tutorial is displayed.