TAKING  pH MEASUREMENTS

• Turn the meter on by connecting the 12 VDC power

adapter to the meter and the mains.

• Connect a pH electrode to the HACCP Indicator.
• Remove the protective cap from the pH electrode and

immerse the tip (4 cm/1½”) in the sample.

• Allow the reading to stabilize.

ADJUSTING THE SETPOINT

You can select a Setpoint between 4.0 and 7.5 pH. A visual

LED indication alerts you when the sample pH is more than

the desired value.
• Unscrew and remove the rear panel and gasket seal to

access the MEASURE/SET switch and the Setpoint ad-

justment trimmer.

• Move the switch up to visualize the Setpoint.

• With a small screwdriver adjust the Setpoint trimmer to

display the desired value. The Setpoint is factory set at

5.5 pH.

• Move the switch down to return to measurement mode.
• Replace the rear panel and the gasket, ensuring the

unit is properly closed.

• Whenever the pH reading ex-

ceeds Setpoint, the alarm LED

blinks to warn the user.

For better accuracy, frequent calibration of the instrument is

recommended. In addition, the instrument must be recali-

brated whenever:

a) The pH electrode is replaced.
b) After testing aggressive samples.
c) Where higher accuracy is required.
d) At least once a month.
e) When the electrode has been used for a long time.

PREPARATION

Pour small quantities of pH 7.0 (HI 7007

HI 7007

HI 7007

HI 7007

HI 7007)  and pH 4.0

(HI 7004

HI 7004

HI 7004

HI 7004

HI 7004)  solution into two clean beakers.

For accurate calibration use two beakers for each buffer

solution, the first one for rinsing the tip of the electrode and

the second one for calibration. This way, contamination of

the buffers is minimized.

A range of HANNA buffer solutions in single-use sachets is

also available for a more hygienic calibration.

CALIBRATION PROCEDURE

• Turn the meter on and make sure that the MEASURE/SET

switch is on the MEASURE mode.

• Remove the protective cap from

the electrode, rinse and im-

merse it in a pH 7.0 buffer.

Stir gently and then wait a

couple of minutes for the read-

ing to stabilize.

Note:

 The electrode should be sub-

merged approximately 4 cm

(1½”) in the solution.

CALIBRATION

CALIBRATION

CALIBRATION

CALIBRATION

CALIBRATION

HI 7004

HI 7007

HI 7004

HI 7007

HI 7004

HI 7007

RINSE

HI 7007

CALIBRATION

• Adjust the right hand trimmer with the calibration

screwdriver until the LCD shows pH 7.0.

Rinse and immerse the pH electrode in

pH 4.0 solution (second buffer) and stir

gently.

• Wait a couple of minutes and then adjust the left hand

trimmer until the LCD shows the value of the second

buffer (4.0 in this case).

The pH calibration is now complete.

4 cm

(1½")

7.0

FOOD STABILITY
p H

I N D I C ATO R

HACCP

pH

pH4

pH7

Recommendations for Users

Before using this product, make sure that it is entirely

suitable for the environment in which it is used. Operation

of this instrument in residential areas could cause unaccept-

able interference to radio and TV equipment.

The glass bulb at the end of the electrode is sensitive to

electrostatic discharge. Avoid touching this glass bulb at all

times. During operation, ESD wrist straps should be worn to

avoid possible damage to the electrode by electrostatic dis-

charge.

Any variation introduced by the user to the supplied equip-

ment may degrade the instrument’s EMC performance.

To avoid electrical shock, do not use this instrument when

voltage at the measurement surface exceeds 24 VAC or 60 VDC.

To avoid damage or burns, do not perform any measurement

in microwave ovens.

A C C E S S O R I E S

A C C E S S O R I E S

A C C E S S O R I E S

A C C E S S O R I E S

A C C E S S O R I E S

FC 200B

Conic open junction pH electrode

FC 230B

pH electrode with optional blade

FC 098

Stainless steel blade, 20mm (0.8”)

FC 099

Stainless steel blade, 35mm (1.4”)

HI 70004P pH 4.01 solution, 20 mL sachet (25 pcs)

HI 70007P pH 7.01 solution, 20 mL sachet (25 pcs)

HI 710005 12 VDC power adapter, US plug

HI 710006 12 VDC power adapter, European plug

HI 710012 12 VDC power adapter, Australian plug

HI 710013 12 VDC power adapter, Southern Africa plug

HI 710014 12 VDC power adapter, UK plug

HI 77400P pH 4 & 7 solutions, 20 mL sachet (5 each)

HI 7004M

pH 4.01 solution, 230 mL

HI 7007M

pH 7.01 solution 230 mL

HI 70300M Storage solution, 230 mL

HI 7061M

Cleaning solution, 230 mL