Hanna Instruments HI 98140 Instruction Manual Download Page 9

17

16

To store the current reading in memory press the LOG key while in
measurement mode. The LCD will display "Stor" along with the "LOG"
indication and the sample number for a few seconds.

By pressing the LOG key a complete set of information is memorized:
date, time, pH and mV (

HI98150

 only) readings, temperature

reading and also a message on the meter status.
Up to 500 samples can be stored into memory.
When the memory is full and the LOG key is
pressed, the sample will not be stored and
the LCD will display "FULL". In this case it is
necessary to delete all the memory to pro-
ceed.

TO VIEW LOGGED DATA

To retrieve the memorized information press ALT and RCL.

The meter displays the date (upper LCD) and
the number (lower LCD) of the last logged
sample. The "ZERO" indication will be dis-
played if no samples are stored in memory.
• Select the desired sample number with the arrow keys. Pressing

the 

 

key while the last sample is displayed causes the meter to

go to the first sample.

• Press RANGE to view remaining data of the

selected sample. After the date information,
the remaining data will be displayed in the
following order:

LOGGING FUNCTION

TEMPERATURE CALIBRATION PROCEDURE

 

(for technical personnel only)

A two point calibration at 0.0, 50.0°C has to be performed in order
to store the new calibration data in memory.
1. Immerse the pH electrode with the built-in temperature sensor in

the 0°C temperature bath.

2. Press ALT and CALT to enter temperature calibration mode.
3. Enter the password.
4. The meter will display "0.0 °C" on the lower LCD with the "BUF 1"

and "CAL" indication.

5. When the reading is stable the "CFM" symbol starts to blink.
6. Press ALT and CFM to confirm. LCD will then display 50.0°C on

the lower LCD with the "BUF 2" indication.

7. Immerse the pH electrode with the built-in temperature sensor in

the 50°C temperature bath.

8. When the reading is stable the "CFM" symbol starts to blink.
9. Press ALT and CFM to confirm and return to normal operational

mode.

10. Set the simulator to +350 mV.
11. When the "CFM" symbol blinks, the read-

ing is stable. Press ALT and CFM to
confirm the second value.

12. The LCD will display "Stor" for a few seconds.
13. If only a two-point calibration is required, press CAL to leave the

calibration mode. Proceed with the next step for three-point
calibration.

14. The meter will display "1900 mV" on the

lower LCD with the "BUF 3" indication.

15. Set the simulator to 1900 mV.
16. When the "CFM" symbol blinks, the read-

ing is stable. Press ALT and CFM to
confirm the third value.

17. Calibration is now complete, the instrument returns to normal

operational mode.

Note:

"WRONG" message notifies the user if the selected value is
wrong.

Summary of Contents for HI 98140

Page 1: ...nual HI 98140 HI 98150 Portable Waterproof pH ORP Meters with Smart Electrodes Log On Demand 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us at www TestEquip...

Page 2: ...so incorporates a temperature sensor elimi nating the need for an additional temperature probe HI 98140 and HI 98150 feature all the most advanced features of laboratory pH meters Moreover they are pr...

Page 3: ...ed data with ALT These instruments also feature a controlled access to calibration and GLP settings through a password protection method The Battery Error Preventing System BEPS detects when the batte...

Page 4: ...re turning the meter off If the meter is OFF detach the electrode from the meter before immersing it in the storage solution ON OFF SPECIFICATIONS Range 4 00 to 19 99 pH 20 to 120 0 C HI 98150 only 60...

Page 5: ...8 10 Calibration alarm time out 01 99 days OFF OFF 20 Auto Off Power down time out ON OFF ON 21 Firmware version 22 Battery level test 99 Password2 0000 9999 0000 Setup can be used to view data regard...

Page 6: ...to change them pressing ALT and FNC with no need to enter setup mode Note If measurements are taken successively in different samples it is recommended to rinse the electrode thoroughly with deion iz...

Page 7: ...om 0000 with the arrow keys 4 Press ALT and CFM to confirm the password or CAL to exit 5 If password is correct the meter displays 7 01pH on the lower LCD with the BUF 1 and CAL indication The upper L...

Page 8: ...be displayed to alert the user 10 If only a single point calibration is re quired press CAL to exit the calibration mode and maintain the previous slope calibration The instrument then checks the ele...

Page 9: ...tion at 0 0 50 0 C has to be performed in order to store the new calibration data in memory 1 Immerse the pH electrode with the built in temperature sensor in the 0 C temperature bath 2 Press ALT and...

Page 10: ...hat an ORP electrode was connected mV reading HI98150 only means reading out of range temperature reading means reading out of range GLP message If RANGE is pressed when the GLP message is displayed t...

Page 11: ...I98140 and HI98150 use electrodes with a built in EEPROM in which calibration data are read at start up and stored after calibration The meter can automatically analyze the data and advise the user if...

Page 12: ...T The LCD can be illuminated to allow the user to see the readings even in dark environments It can be enabled disabled through the LIGHT key and it is automatically disabled when batter ies are in lo...

Page 13: ...charged batteries The backlight can be enabled 2 Weakening batteries The backlight is automatically disabled and it is not possible to enable it until new batteries are inserted or an external power...

Page 14: ...tuate up and down could be due to Clogged Dirty Junction refer to the cleaning procedure above Loss of shielding due to low electrolyte level for refillable electrodes only refill with fresh HI 7071 f...

Page 15: ...ctrode Slow Response Excessive Drift soak the tip in HI 7061 solution for 30 minutes rinse thoroughly in distilled water and then follow the Cleaning Procedure above SMART ELECTRODES HI 1615D Smart pH...

Page 16: ...rights are reserved Reproduction in whole or in part is prohibited without the written consent of the copyright owner Hanna Instruments reserves the right to modify the design construction and appeara...

Reviews: