INITIAL PREPARATION

Each meter is supplied complete with batteries. Remove the back
cover, unwrap the batteries and install them while paying attention
to the polarity. Alternatively, connect the supplied 12 Vdc adapter to
the power adapter plug.
To prepare the instrument for use, connect the pH or ORP (

HI 98150

only) electrode to the connector located on the top of the instrument.

To switch the meter on, press the ON/OFF key.
The batteries charge status or "LINE" message
(if external power adapter is connected) will be
displayed on the LCD for a few seconds.
The meter is now ready to operate.

To maximize battery life, the meter is automatically switched off after 5
minutes of non-use (this feature is enabled by default and can be
disabled through setup code 20). To reactivate the instrument press
the ON/OFF key.
Before proceeding with pH measurements follow the calibration proce-
dure.

Note:

When the use of an alternate function (DEL, FNC, CFM, GLP,
RCL and CALT) is requested, press and hold the ALT key first
and then the second listed key.

Note:

 to prevent damage to the electrode, remove it from the sample

before turning the meter off.

If the meter is OFF, detach the electrode from the meter before
immersing it in the storage solution.

ON/OFF

SPECIFICATIONS

Range

-4.00 to 19.99 pH / -20 to 120.0°C

HI 98150 

only: ±600.0 mV; ±2000 mV

Resolution

0.01 pH

0.1°C (-10 to 120°C) / 1°C (below -10°C)

HI 98150 

only: 0.1 mV (±400 mV) /

0.2 mV (±400 to ±600 mV) / 1 mV (outside)

Accuracy (@20°C/68°F)

±0.01 pH

±0.4 °C from 0 to 70°C / ±1°C outside

HI 98150 

only: ±0.2 mV (±400 mV) /

±0.4 mV (±400 to ±600 mV) / ±2 mV (outside)

Typical EMC Deviation

±0.02 pH / ±0.4°C

HI 98150 

only: ±1 mV (±600 mV) / ±2 mV (outside)

pH Calibration

Automatic, 1 or 2 point with 5 memorized buffers

(pH 4.01, 6.86, 7.01, 9.18 and 10.01)

mV Calibration

Automatic, 2 point at 0 and 350 mV or

(

HI98150 

only)

3 point at 0, 350 and 1900 mV

Temperature Compensation

Automatic, -20 to 120°C (-4 to 248°F)

pH Electrode

HI 1618D

 smart electrode with built-in temperature

sensor, DIN connector and 1 m (3.3') cable (included)

ORP Electrode

Smart ORP electrode with DIN connector

(

HI98150

 only; see accessories)

Input Impedance

10

12 

Ohm

PC Connection

Through RS232 serial port,

using 

HI 920011

 connection cable & 

HI 92000

 software

Power Supply

4 x 1.5V AA batteries / approx. 300 hours of use

(with backlight off) or 12 Vdc input

Casing

IP 67

Environment

0 to 50°C (32 to 122°F); RH max 100%

Dimensions

196 x 80 x 60 mm (7.7 x 3.1 x 2.4")

Weight

500 g (1.1 lb.)

Summary of Contents for HI 98140

Page 1: ...nual HI 98140 HI 98150 Portable Waterproof pH ORP Meters with Smart Electrodes Log On Demand 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us at www TestEquip...

Page 2: ...so incorporates a temperature sensor elimi nating the need for an additional temperature probe HI 98140 and HI 98150 feature all the most advanced features of laboratory pH meters Moreover they are pr...

Page 3: ...ed data with ALT These instruments also feature a controlled access to calibration and GLP settings through a password protection method The Battery Error Preventing System BEPS detects when the batte...

Page 4: ...re turning the meter off If the meter is OFF detach the electrode from the meter before immersing it in the storage solution ON OFF SPECIFICATIONS Range 4 00 to 19 99 pH 20 to 120 0 C HI 98150 only 60...

Page 5: ...8 10 Calibration alarm time out 01 99 days OFF OFF 20 Auto Off Power down time out ON OFF ON 21 Firmware version 22 Battery level test 99 Password2 0000 9999 0000 Setup can be used to view data regard...

Page 6: ...to change them pressing ALT and FNC with no need to enter setup mode Note If measurements are taken successively in different samples it is recommended to rinse the electrode thoroughly with deion iz...

Page 7: ...om 0000 with the arrow keys 4 Press ALT and CFM to confirm the password or CAL to exit 5 If password is correct the meter displays 7 01pH on the lower LCD with the BUF 1 and CAL indication The upper L...

Page 8: ...be displayed to alert the user 10 If only a single point calibration is re quired press CAL to exit the calibration mode and maintain the previous slope calibration The instrument then checks the ele...

Page 9: ...tion at 0 0 50 0 C has to be performed in order to store the new calibration data in memory 1 Immerse the pH electrode with the built in temperature sensor in the 0 C temperature bath 2 Press ALT and...

Page 10: ...hat an ORP electrode was connected mV reading HI98150 only means reading out of range temperature reading means reading out of range GLP message If RANGE is pressed when the GLP message is displayed t...

Page 11: ...I98140 and HI98150 use electrodes with a built in EEPROM in which calibration data are read at start up and stored after calibration The meter can automatically analyze the data and advise the user if...

Page 12: ...T The LCD can be illuminated to allow the user to see the readings even in dark environments It can be enabled disabled through the LIGHT key and it is automatically disabled when batter ies are in lo...

Page 13: ...charged batteries The backlight can be enabled 2 Weakening batteries The backlight is automatically disabled and it is not possible to enable it until new batteries are inserted or an external power...

Page 14: ...tuate up and down could be due to Clogged Dirty Junction refer to the cleaning procedure above Loss of shielding due to low electrolyte level for refillable electrodes only refill with fresh HI 7071 f...

Page 15: ...ctrode Slow Response Excessive Drift soak the tip in HI 7061 solution for 30 minutes rinse thoroughly in distilled water and then follow the Cleaning Procedure above SMART ELECTRODES HI 1615D Smart pH...