Hanna Instruments HI 98140 Instruction Manual Download Page 14

27

26

Replace solution in the protective cap with a few drops of 

HI 70300

storage solution or, in its absence, filling solution (

HI 7071 

for single

junction or 

HI 7082 

for double junction electrodes). Follow the

Preparation Procedure above before taking measurements.

Note:

NEVER STORE THE ELECTRODE IN DISTILLED WATER OR DRY.

PERIODIC MAINTENANCE

Inspect the electrode and the cable. The cable used for connection to
the meter must be intact and there must be no points of broken
insulation on the cable or cracks on the electrode stem or bulb.
Connectors must be perfectly clean and dry.
If any scratches or cracks are present, replace the electrode.
Rinse off any salt deposits with water.

For refillable electrodes

Refill it with fresh electrolyte (

HI 7071

 for single junction or 

HI 7082

for double junction electrodes). Allow the electrode to stand upright for
1 hour. Follow the Storage Procedure above.

CLEANING PROCEDURE

- General

:

soak in HANNA 

HI 7061

 general cleaning solu-

tion for approximately 1 hour.

Removal of films, dirt or deposits on the membrane/junction:

Protein

:

soak in HANNA 

HI 7073

 protein cleaning solution

for 15 minutes

Inorganic:

soak in Hanna 

HI 7074

 inorganic cleaning solu-

tion for 15 minutes

Oil/grease:

rinse with Hanna 

HI 7077

 Oil & Fat cleaning

solution

IMPORTANT:

 After performing any of the cleaning procedures, rinse

the electrode thoroughly with distilled water, drain and refill the
reference chamber with fresh electrolyte (not necessary for gel filled
electrodes), and soak the electrode in 

HI 70300

 storage solution for at

least 1 hour before taking measurements.

TROUBLESHOOTING

Evaluate performance of your electrode based on the following
possibilities.

• Noise 

(readings fluctuate up and down) could be due to:

- Clogged/Dirty Junction: 

refer to the cleaning procedure above.

- Loss of shielding

 due to low electrolyte level (for refillable

electrodes only): refill with fresh 

HI 7071 

for single junction

electrodes or 

HI 7082 

for double junction electrodes.

Note:

 To prevent damage to the pH electrode, remove it from the

sample before turning the meter off.

With the meter OFF, detach the electrode from the meter and
immerse it in the storage solution.

PREPARATION

Remove the protective cap.
DO NOT BE ALARMED IF SALT DEPOSITS ARE PRESENT.
This is normal with electrodes and they will disappear when rinsed
with water.
During transport tiny bubbles of air may form inside the glass bulb.
The electrode cannot function properly under these conditions. These
bubbles can be removed by "shaking down" the electrode as you
would do with a glass thermometer.
If the bulb and/or junction are dry, soak the electrode in 

HI 70300

storage solution for at least one hour.

For refillable electrodes

If  the filling solution (electrolyte) is less than 1 cm (½") below the fill
hole, add 

HI 7082

 (3.5M KCl electrolyte solution) for double junction

or 

HI

 

7071

 (3.5M KCl+AgCl electrolyte solution) for single junction

electrodes.
For a faster response unscrew the fill hole screw during measurements.

For AmpHel

®

 electrodes

If the electrode does not respond to pH changes, the battery is run
down and the electrode should be replaced.

MEASUREMENT

Rinse the electrode tip with distilled water.
Immerse the tip (4 cm /1½") in the sample and stir gently for
approx. 30 seconds.
For a faster response and to avoid cross contamination of the samples,
rinse the electrode tip with a few drops of the solution to be tested,
before taking measurements.

STORAGE

To minimize clogging and ensure a quick response time, the glass
bulb and the junction should be kept moist and not allowed to dry
out.

ELECTRODE CONDITIONING

AND MAINTENANCE

Summary of Contents for HI 98140

Page 1: ...nual HI 98140 HI 98150 Portable Waterproof pH ORP Meters with Smart Electrodes Log On Demand 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us at www TestEquip...

Page 2: ...so incorporates a temperature sensor elimi nating the need for an additional temperature probe HI 98140 and HI 98150 feature all the most advanced features of laboratory pH meters Moreover they are pr...

Page 3: ...ed data with ALT These instruments also feature a controlled access to calibration and GLP settings through a password protection method The Battery Error Preventing System BEPS detects when the batte...

Page 4: ...re turning the meter off If the meter is OFF detach the electrode from the meter before immersing it in the storage solution ON OFF SPECIFICATIONS Range 4 00 to 19 99 pH 20 to 120 0 C HI 98150 only 60...

Page 5: ...8 10 Calibration alarm time out 01 99 days OFF OFF 20 Auto Off Power down time out ON OFF ON 21 Firmware version 22 Battery level test 99 Password2 0000 9999 0000 Setup can be used to view data regard...

Page 6: ...to change them pressing ALT and FNC with no need to enter setup mode Note If measurements are taken successively in different samples it is recommended to rinse the electrode thoroughly with deion iz...

Page 7: ...om 0000 with the arrow keys 4 Press ALT and CFM to confirm the password or CAL to exit 5 If password is correct the meter displays 7 01pH on the lower LCD with the BUF 1 and CAL indication The upper L...

Page 8: ...be displayed to alert the user 10 If only a single point calibration is re quired press CAL to exit the calibration mode and maintain the previous slope calibration The instrument then checks the ele...

Page 9: ...tion at 0 0 50 0 C has to be performed in order to store the new calibration data in memory 1 Immerse the pH electrode with the built in temperature sensor in the 0 C temperature bath 2 Press ALT and...

Page 10: ...hat an ORP electrode was connected mV reading HI98150 only means reading out of range temperature reading means reading out of range GLP message If RANGE is pressed when the GLP message is displayed t...

Page 11: ...I98140 and HI98150 use electrodes with a built in EEPROM in which calibration data are read at start up and stored after calibration The meter can automatically analyze the data and advise the user if...

Page 12: ...T The LCD can be illuminated to allow the user to see the readings even in dark environments It can be enabled disabled through the LIGHT key and it is automatically disabled when batter ies are in lo...

Page 13: ...charged batteries The backlight can be enabled 2 Weakening batteries The backlight is automatically disabled and it is not possible to enable it until new batteries are inserted or an external power...

Page 14: ...tuate up and down could be due to Clogged Dirty Junction refer to the cleaning procedure above Loss of shielding due to low electrolyte level for refillable electrodes only refill with fresh HI 7071 f...

Page 15: ...ctrode Slow Response Excessive Drift soak the tip in HI 7061 solution for 30 minutes rinse thoroughly in distilled water and then follow the Cleaning Procedure above SMART ELECTRODES HI 1615D Smart pH...

Page 16: ...rights are reserved Reproduction in whole or in part is prohibited without the written consent of the copyright owner Hanna Instruments reserves the right to modify the design construction and appeara...

Reviews: