6
SPECIFICA
TIONS
Range
0.00 to 3.50 mg/L (as SDBS)
Resolution
0.01 mg/L
Accuracy
@25
°
C (77
°
F)
±0.04 mg/L ±3 % of reading
Light Source
Tungsten lamp
Light Detector
Silicon Photocell with narrow band interference filter @610 nm
Method
Adaptation of the USEPA method 425.1 and Standard Methods for the
Examination of Water and Wastewater, 20
th
edition, 5540C, Anionic Sur‑
factants as MBAS.
Environment
0 to 50
°
C (32 to 122
°
F)
max 95 % RH non‑condensing
Battery Type
9V (1 pc.)
Auto‑Shut off
After 10’ of non‑use in measurement mode; after 1 hour of non‑use in
calibration mode; with last reading reminder
Dimensions
192 x 104 x 69 mm (7.6 x 4.1 x 2.7”)
Weight
320 g (11.3 oz.)
REQUIRED REAGENTS
Code
Description
Quantity/test
HI95769A‑0
Anionic Surfactants Reagent A 4 drops
HI95769B‑0
Anionic Surfactants Reagent B 2 drops
Chloroform
10 mL
Dionized Water
15 mL