9

8

HI935009

HI935005 / 5005N

OPERATIONAL GUIDE

To switch the instrument ON, press the ON/OFF key. 
The thermometer will carry out a self diagnostic test routine, the LCD will 
show all segments for a few seconds (or as long as ON/OFF is held), 
followed by the percentage indication of the remaining battery life.
The thermometer then enters normal measurement mode. 

If a temperature probe is plugged in, the meter displays the measured 
temperature. 
If no probe is plugged in, or if the reading is over-range, the display 
shows flashing dashes. 
If a measurement is slightly over the range of 
the meter specifications, the display will flash 
the closest full-scale value.
To switch the meter OFF, press the ON/OFF key.

°C/°F SELECTION

Measurements can be displayed in either degrees Celsius or Fahrenheit. 
The meter is factory set to °C scale; to change the scale, press °C/°F 
key or (ALT+) °C/°F keys, or set the switch located in the battery 
compartment (

HI935002 

only, see figure on page 10). 

HOLD and MR FUNCTIONS

The HOLD function is activated by pressing the HOLD key. 
The measured temperature is held on the display until HOLD is pressed 
again. The "HOLD" tag blinks on the display while in HOLD mode.

Note

: Although the display is frozen, internally the meter continues 

measuring and updating Hi/Lo and relative values.

In 

HI935009

 only, the held value is also stored in non-volatile memory 

and can be recalled by pressing (ALT+) MR key. While the MR key 
is held, the "HOLD" tag lights up and the meter displays the stored 
value; when the MR key is released, after 1 second the meter returns 
to normal measurement mode.

HIGH/LOW TEMPERATURES

The maximum and minimum temperatures 
are continuously monitored and displayed 
on the lower portion of the LCD. 

Note

: When the reading goes over-range or the probe is removed, 

the Hi and Lo values display dashes until cleared. 

CLEAR FUNCTION

Upon pressing the CLR key, the High/Low values may be cleared at any 
time during measurement and the current reading is assigned to the 
highest and lowest temperature values for the displayed channel only.

AUTO SHUT-OFF (HI935005 and HI935009)

To save battery life, the 

HI935005

 and 

HI935009

 models are provided 

with an auto-off feature, which switches the instrument off after a 
certain period of nonuse.
The delay time can be changed by the user (8 minutes, 60 minutes, 
or disabled). 
To access this feature, the 

HI935005

 is provided with an internal button 

located in the battery compartment (see figure on page 10). Unscrew 
the 4 screws on the back of the meter and select the desired auto-off 
time period by pressing the button and viewing the value on the LCD. 
Re-attach the back making sure that the gasket is in place and tighten 
the 4 screws to ensure a watertight seal. 

Summary of Contents for HI 935005

Page 1: ...Instruction Manual HI935005 HI935005N HI935002 HI935009 Portable K Thermocouple Thermometers www hannainst com...

Page 2: ...hich have been designed using the latest microprocessor technology to provide reliable and accurate measurements in a wide temperature range Standard features include K type probes waterproof casing d...

Page 3: ...ar the HI and LO values BL HI935005N only toggle the backlight ON and OFF CAL HI935005N only press and hold for 5 seconds to enter the Calibration mode with reading within 3 C range press to abort the...

Page 4: ...ment DISP select Normal Relative or T1 T2 measuring mode ALT HI935009 only enable the second function keys the ALT tag turns on to indicate that the second functions are enabled Note The ALT key can b...

Page 5: ...updating Hi Lo and relative values In HI935009 only the held value is also stored in non volatile memory and can be recalled by pressing ALT MR key While the MR key is held the HOLD tag lights up and...

Page 6: ...corresponding tags will light up to inform the user To select the desired channel use the T1 T2 key The 2 channel models are also provided with the DISP function which allows the user to select the i...

Page 7: ...e made from distilled water Immerse the temperature probe in the center of the ice bath taking care not to touch the ice with the probe tip Ensure that the meter is measuring a temperature within 3 C...

Page 8: ...ction with the HI766HD probe handle HI766PA Roller surface probe max 320 C 600 F HI766PB Surface probe max 650 C 1200 F HI766PC Penetration probe max 900 C 1650 F HI766PD Air probe max 300 C 570 F HI7...

Page 9: ...ruments Inc Highland Industrial Park 584 Park East Drive Woonsocket RI 02895 USA www hannainst com Local Office Hanna Instruments Inc Highland Industrial Park 584 Park East Drive Woonsocket RI 02895 U...

Reviews: