Dear Customer,

Thank you for choosing a HANNA product.

Please read this instruction manual carefully
before using the instrument.

This manual will provide you with the neces-
sary information for a correct use of the instru-
ment.

If you need additional technical support, do not
hesitate to e-mail us at [email protected]

These instruments are in compliance with the
CE directives.

Preliminary Examination ............................. 3
General Description ..................................... 3
Functional Description - Probe .................... 4
Functional Description - Meter ..................... 5
Specifications ............................................. 6
Probe Preparation ....................................... 7
Calibration ................................................... 8
Taking Measurements ............................... 10
Altitude Compensation .............................. 12
Salinity Compensation .............................. 13
Probe & Membrane Maintenance .............. 14
Battery Replacement ................................ 16
Accessories .............................................. 17
Warranty ................................................... 18
CE Declaration of Conformity .................... 19

PRELIMINARY EXAMINATION

Remove the instrument from the packing material
and examine it to make sure that no damage
has occurred during shipping. If there is any
damage, immediately notify your dealer.

Each meter is supplied complete with:

HI 76407/4 DO probe with 4 m (13’) cable

2 spare membranes with O-rings

Protective cap

HI 7041S electrolyte solution (30 mL)

Batteries (4 x 1.5V AA)

Instruction manual

Rugged carrying case

Note: Save all packing material until you are

sure that the instrument functions
correctly. Any defective items must be
returned in the original packing with the
supplied accessories.

GENERAL  DESCRIPTION

HI 9143 and HI 9145 are water-resistant
dissolved oxygen meters 

designed for harsh

outdoor applications, especially waste water
treatment and fish farming.

Simple on-site calibration requires no chemical
solutions. Just expose the probe to air and
press the CAL button. In a few minutes, the
meter is calibrated and ready to use.

The new and improved HI 9143 and HI 9145
measure and display O

2

 from 0 to 300%, 0 to

45 mg/L and temperature from 0 to 50 °C.

In addition to the features of HI 9145HI 9143
also supplies the user with compensation of
salinity and altitude factors.

Both meters compensate for the temperature
effect.

TABLE OF CONTENTS

Summary of Contents for HI 9143

Page 1: ...ruction Manual HI 9143 HI 9145 Portable D O Meters in Water Resistant Casing HI 9143 Microprocessor Auto Cal Dissolved Oxygen Meter CAL ON OFF RANGE FACTOR SAMPLE HOLD ppm C w w w h a n n a i n s t c...

Page 2: ...mediately notify your dealer Each meter is supplied complete with HI76407 4 DOprobewith4 m 13 cable 2 spare membranes with O rings Protectivecap HI 7041S electrolyte solution 30 mL Batteries 4 x 1 5V...

Page 3: ...et for 12 Vdc power adapter 19 LOW BAT low battery indicator 20 SAMPLE measuring mode indicator 21 F1 altitude factor indicator 22 F2 salinity factor indicator 23 or ppm oxygen measure unit 24 Seconda...

Page 4: ...the cap clockwise Some electrolyte will over flow When not in use protect the membrane with the supplied cap SPECIFICATIONS Shipping cap black red THEN SCREW BACK ON THIRD TAP SECOND FILL FIRST INSER...

Page 5: ...table measurements with the same recurring degree of accuracy With the probe properly polarized oxygen is continually consumed bypassingthroughthe sensitivediaphragmanddissolvingintheelec trolyte solu...

Page 6: ...e ambient temperature and the temperatureofthesample thelongeritwilltake the probe to acclimatized itself to the sample TAKINGMEASUREMENTS Make sure the meter has been calibrated and the protective ca...

Page 7: ...1 8 7 8 3 8 0 7 7 7 3 7 1 6 8 6 6 6 3 6 1 6 0 5 7 5 6 5 4 1800 m 11 8 11 2 10 6 10 1 9 6 9 2 8 8 8 3 8 0 7 7 7 4 7 1 6 8 6 6 6 3 6 1 5 9 5 8 5 5 5 4 5 2 F 32 0 35 6 39 2 42 8 46 4 50 0 53 6 57 2 60 8...

Page 8: ...tilled or deionized water If any imperfection is observed or any damage is evident such as wrinkles tears or holes themembraneshouldbereplaced Make sure that the O Ring is properly seated in the membr...

Page 9: ...on the top of the instrument and plug the power adapter into the socket Note The instrument uses the following con figuration Use the HANNA adapters such as HI 710005 orHI710006 withproperpolarityconf...

Page 10: ...rding to the instruc tions Probes are warranted for six months Thiswarrantyislimitedtorepairorreplacement freeofcharge Damagesduetoaccidents misuse tampering or lack of prescribed maintenance are not...

Page 11: ...3 3258 9567 Korea Tel 02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 SouthAfrica Tel 011 615 6076 Fax 011 615 8582 Taiwan Tel 886 2 27...

Reviews: